RESERVOIR "Y" ADAPTER, (a) BT920 (b) BT926 Suction and Anticoagulation assemblies are indicated for use with...
Medtronic Perfusion Systems began a recall of RESERVOIR "Y" ADAPTER, (a) BT920 (b) BT926 Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip... on . FDA classified it as Class II and gives the reason as "Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0785-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Perfusion Systems
- Product
- RESERVOIR "Y" ADAPTER, (a) BT920 (b) BT926 Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip and reservoir.
- Product code
- CAC Apparatus, Autotransfusion
- Reason for recall
- Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm, Medtronic, began notifying US consignees on 27-Nov-2018 via hand delivery to be followed up with a UPS 2-Day communication titled "URGENT MEDICAL DEVICE RECALL" to each listed consignee that does not confirm hand delivery. Communications OUS also began 27-Nov-2018 using locally approved methods. The letter described the product, problem and actions to be taken. Customers were requested to do the following: 1. Identify and quarantine all unused affected product as listed in your inventory. 2. Return all unused affected product in your inventory to Medtronic. Contact Medtronic Customer Service at 1-800-854-3570 to initiate a product return. Your local Medtronic Representative can assist you in the return and replacement of this product as necessary. 3. Complete the enclosed Customer Confirmation Certificate and email to [email protected]. If you have any questions about this action, please contact your Medtronic representative or call Director, Cardiac and Vascular Group, at 763-526-2513 or email: [email protected].
- Quantity in commerce
- 1100 units
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, France, Germany, Japan, and Sweden.
- Product codes and lots
- BT920: GTIN/UDI Lot/Serial Numbers 10673978522460 216044420 10673978522460 216062719 10673978522460 216438180 BT926: GTIN/UDI Lot/Serial Numbers 10673978522477 215782209 10673978522477 215903433 10673978522477 216034158 10673978522477 216107790 10673978522477 216372963 10673978522477 216438301
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0785-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0786-2019 | Class II | Ongoing | Voluntary: Firm initiated |