Z-0352-2020Class II
ProtectIV Plus Safety IV Catheter, sterile. Packaged in a pouch and than into an inner carton
Abbott Medical / initiated 2018-11-26 / Terminated / root cause: manufacturing process
Smiths Medical ASD Inc. began a recall of ProtectIV Plus Safety IV Catheter, sterile. Packaged in a pouch and than into an inner carton on . FDA classified it as Class II and gives the reason as "Certain lots of the Model 306701 ProtectIV Plus Safety IV Catheter may exhibit damage on the safety mechanism guard". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0352-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ProtectIV Plus Safety IV Catheter, sterile. Packaged in a pouch and than into an inner carton.
- Product code
- FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
- Reason for recall
- Certain lots of the Model 306701 ProtectIV Plus Safety IV Catheter may exhibit damage on the safety mechanism guard.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- On about November 26, 2018, Smiths Medical sent an URGENT MEDICAL DEVICE RECALL NOTICE via email or FedEx, (where email addresses are not available) and accompanying Recall Response Forms to all consignees who purchased potentially affected product. Consignees with potentially affected product were instructed to immediately segregate the product and return them to Stericycle, via pre-paid shipping label. They we also instructed to notify customers if the affected product was further distributed and complete and return the Recall Response Form. Questions can be submitted to Smith Medical via email at [email protected].
- Quantity in commerce
- 219,150 devices
- Distribution
- United States - AK, AZ, FL, GA, IL, IN, KS, LA, MD, MN, MS, NC, NJ, NY, OH, PA, SC, TN, TX, VA, and WA. One foeign consignee in Costa Rica
- Product codes and lots
- Model number 306701; Lot numbers: 3635765, 3639446, 3642834, 3646828, 3646829, 3646830, 3646831, 3646832, 3646833, 3651082.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0352-2020 | Class II | Terminated | Voluntary: Firm initiated |