Z-0629-2019Class II

Jarvik 2000 Ventricular Assist System external components: JHI-212 Y Cable, JHI-202 Lithium-ion Battery...

Jarvik Heart Inc / initiated 2018-10-15 / Terminated / root cause: design

Jarvik Heart Inc began a recall of Jarvik 2000 Ventricular Assist System external components: JHI-212 Y Cable, JHI-202 Lithium-ion Battery Cable, JHI-204 Extensions Cable, JHI-201 Pedestal Cable on . FDA classified it as Class II and gives the reason as "There is a potential for the male connectors on external cables to become damaged. This could include a detached male connector housing, misaligned connector, cracked connector housing, and bent pins". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0629-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Jarvik Heart Inc
Product
Jarvik 2000 Ventricular Assist System external components: JHI-212 Y Cable, JHI-202 Lithium-ion Battery Cable, JHI-204 Extensions Cable, JHI-201 Pedestal Cable
Product code
DSQ Ventricular (Assist) Bypass
Reason for recall
There is a potential for the male connectors on external cables to become damaged. This could include a detached male connector housing, misaligned connector, cracked connector housing, and bent pins.
Root cause tag
design
FDA root cause
Device Design
Action
Urgent Field Safety Notice letters dated 10/5/18 were distributed to customers.
Quantity in commerce
274
Distribution
Products were distributed to the following US states: FL, NY, and UT. Products were distributed to the following foreign countries: Croatia, Czech Republic, France, Hungary, India, Italy, and Japan.
Product codes and lots
All systems/cables

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0629-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0629-2019Class IITerminatedVoluntary: Firm initiated