Z-0704-2019Class II

VariCam

GE Healthcare / initiated 2017-10-30 / Completed

GE Healthcare, LLC began a recall of VariCam on . FDA classified it as Class II and gives the reason as "A potential hazardous situation may occur if the collimator locking handle is not properly placed in the lock position and the warning circuit does not detect that the collimator is unlocked because of the patient...". As of , the recall status on record is Completed.

Recall record

Recall number
Z-0704-2019
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
GE Healthcare
Product
VariCam
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
A potential hazardous situation may occur if the collimator locking handle is not properly placed in the lock position and the warning circuit does not detect that the collimator is unlocked because of the patient proximity to the collimator during a clinical scan.
FDA root cause
Other
Action
The firm initiated the medical device correction by letter on 10/0/2017. The letter instructed the consignee to ensure the collimator locking handles are in the locked position at all times. GE Healthcare will inspect and if necessary correct all affected products.
Quantity in commerce
64 units
Distribution
Worldwide
Product codes and lots
Serial Numbers: 0000000010-203 0000000010-191 000003100-1829 00008125-18-18 00000000010-15 00000000010-63 00008125-16-59 000000099-4-04 00000000010-31 00000000010-82 00008125-16-70 00008125-18-09 00008125-18-16 00000000017-64 10280 00000000010-30 000003681-1803 00000000018-24 000003100-1840 000003101-1502 000003100-1744 00000000017-36 00000000016-86 00000000017-31 1065 1998 1999 000008125-1630 1008 00000000017-20 00000000017-22 00000000017-23 1804 00000000030-14 00000000030-15 00000000010-37 00000000010-44 00000000010-55 00000000030-16 00000000018-39 1704 00000ICPO50612 1697 1696 000000163-1758 00000000018-23 00000000016-73 00000000030-04 00000000010-59 00000000018-22 000003100-1760 000003100-1743 0000000010-134 000003100-1719 343 000003100-1763 61061 00000000010-04 00000000030-11 1649 00000000010-67 00000000010-45 00000000017-27 00000000017-15

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0704-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0703-2019Class IICompletedVoluntary: Firm initiated
Z-0704-2019Class IICompletedVoluntary: Firm initiated