Z-0850-2019Class II

Kaguya Automated Peritoneal Dialysis System

Baxter Healthcare Corporation / initiated 2018-12-11 / Terminated / root cause: software

Baxter Healthcare Corporation began a recall of Kaguya Automated Peritoneal Dialysis System Product Usage: It is intended for automatic control of dialysate solution exchanges in the treatment of adult renal... on . FDA classified it as Class II and gives the reason as "Potential for the software on Automated PD System cyclers which can cause shortened dwell times during a Cycle-Based therapy". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0850-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Kaguya Automated Peritoneal Dialysis System Product Usage: It is intended for automatic control of dialysate solution exchanges in the treatment of adult renal failure patients undergoing Peritoneal dialysis.
Product code
FKX System, Peritoneal, Automatic Delivery
Reason for recall
Potential for the software on Automated PD System cyclers which can cause shortened dwell times during a Cycle-Based therapy
Root cause tag
software
FDA root cause
Software design
Action
Baxter Healthcare Corporation notified providers on about 12/11/2018 of the recall concerning the AMIA Automated Peritoneal Dialysis System, via "Urgent Medical Device Correction" letter. Patients were notified on about 1/14/2018 via "Urgent Medical Device Correction" letter. Providers were instructed to continue using the AMIA Automated PD System however, for patients who are prescribed a Cycle-Based therapy, to change these therapies to Time-Based by following the programming instructions within the Clinician Guide, Section 5.3.1 - Time-Based Programming Method. In Time-Based therapy, to maintain a fixed dwell time, the Smart Dwell treatment option needs to be Disabled. To disable the Smart Dwell treatment option, follow the programming instructions within the Clinician Guide, Section 5.3.1 - Time-Based Programming Method. Questions can be directed to Baxter's Renal Clinical Helpline at 888-736-2543, option 2, Monday through Friday, 8:00 am to 4:30 pm (CT). Providers were informed that Baxter would contact them when the software update is available and provide further instructions on how to download the upgrade remotely. They were also instructed to complete and return the Baxter Customer Reply Form and to inform customers if the device was further distributed to other facilities or departments within their institution. Patients were instructed to continue using the AMIA Automated PD System however, they were directed to check the Programming Method to ensure they have not been prescribed Cycle-Based therapy. The programming method can be verified by following section 10.3 of the Patient Guide: 1. touch "GO" on the home screen and the Confirm Treatment screen should appear, 2. touch "Program Settings" to view the program settings details and the Program Details screen will appear, 3. display the "Program Method" by navigating to Page 3 by pressing the down arrow at the bottom of the screen. If the Programming Method is Time-Based or Tidal, the iss
Quantity in commerce
1,085 devices
Distribution
Worldwide Distribution - US Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Distributed internationally to Canada and Japan.
Product codes and lots
Product code T5C8500. All serial numbers are affected.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0850-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0849-2019Class IITerminatedVoluntary: Firm initiated
Z-0850-2019Class IITerminatedVoluntary: Firm initiated