ICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 60" Tubing...
ICU Medical, Inc. began a recall of ICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 60" Tubing, Disposable Transducer, 3 ml Intraflo Flush, Macrodrip (Pole... on . FDA classified it as Class II and gives the reason as "Reports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay in therapy, and in rare circumstances...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1072-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 60" Tubing, Disposable Transducer, 3 ml Intraflo Flush, Macrodrip (Pole Mount), REF: 42641-06, Sterile EO
- Product code
- DRS Transducer, Blood-Pressure, Extravascular
- Reason for recall
- Reports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay in therapy, and in rare circumstances, pressure tubing separations could lead to blood loss, air embolism, or contamination of the fluid path.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- On 11/27/18, Urgent Medical Device Recall notices were sent via traceable mailing to customers informing them to discontinue use and distribution of the affected product. In addition, customers were asked to quarantine and return affected product, and to complete and return response forms. Customers were instructed to inform all potential users of the product within their organization, and if the product had been further distributed, to immediately notify these additional customers. Customers with additional questions were encouraged to call: Global Complaint Management @ 1-844-654-7780 ([email protected]), Medical Information @ 1-800-241-4002 option 6 ([email protected]), or Customer Care @ 1-949-366-4208.
- Distribution
- U.S.: CA, TX, WI, NY, NC, GA, PA, and NY. Foreign: Canada
- Product codes and lots
- Lot/UDI: 3529857/ (01)0 0840619 04495 4 (17) 200901 (30) 01 (10) 3529857
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1070-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1071-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1072-2019 | Class II | Terminated | Voluntary: Firm initiated |