Heating Profile PRISMAFLO II S blood return warmer for Prismaflex
Stihler Electronic Gmbh began a recall of Heating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP33 Product Usage: The PRISMAFLO IIS warmer is used for warming return... on . FDA classified it as Class II and gives the reason as "PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0759-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stihler Electronic GmbH
- Product
- Heating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP33 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.
- Product code
- KOC Accessories, Blood Circuit, Hemodialysis
- Reason for recall
- PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- STIHLER ELECTRONIC GMBH notified distributors by Registered mail/email and to Head of Biomedical on 09/06/18 identifying the problem, health risk and action to be taken: identify the Heating Profiles according to the serial numbers specified. These profiles must be replaced based on your accurate reply. Complete Inventory must be accounted for. For your reply use attached "Return Form" to Stihler Electronic GmbH. Send the filled return form via e-mail to nicola. vosseler@sti hlerelectronic. de. 3. You will receive the replacement Heating Profiles from our distributor Futuremed America Inc., 15700 Devonshire Street, Granada Hills, 91344 California, US.
- Quantity in commerce
- 821
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA and Foreign of: Switzerland
- Product codes and lots
- Serial Numbers: E22000 to E23999
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0759-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0760-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0761-2019 | Class II | Terminated | Voluntary: Firm initiated |