Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0761-2019Class II

Heating Profile PRISMAFLO II blood return warmer for Prismaflex

Stihler Electronic GmbH / initiated 2018-09-06 / Terminated / root cause: design

Stihler Electronic Gmbh began a recall of Heating Profile PRISMAFLO II blood return warmer for Prismaflex Product Code: PF2-WP2618 Product Usage: The PRISMAFLO IIS warmer is used for warming return... on . FDA classified it as Class II and gives the reason as "PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0761-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stihler Electronic GmbH
Product
Heating Profile PRISMAFLO II blood return warmer for Prismaflex Product Code: PF2-WP2618 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.
Product code
KOC Accessories, Blood Circuit, Hemodialysis
Reason for recall
PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
Root cause tag
design
FDA root cause
Device Design
Action
STIHLER ELECTRONIC GMBH notified distributors by Registered mail/email and to Head of Biomedical on 09/06/18 identifying the problem, health risk and action to be taken: identify the Heating Profiles according to the serial numbers specified. These profiles must be replaced based on your accurate reply. Complete Inventory must be accounted for. For your reply use attached "Return Form" to Stihler Electronic GmbH. Send the filled return form via e-mail to nicola. vosseler@sti hlerelectronic. de. 3. You will receive the replacement Heating Profiles from our distributor Futuremed America Inc., 15700 Devonshire Street, Granada Hills, 91344 California, US.
Quantity in commerce
247 units
Distribution
Worldwide Distribution - US Nationwide in the states of CA and Foreign of: Switzerland
Product codes and lots
Serial Numbers: UH07154 to UH07479

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0761-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0759-2019Class IITerminatedVoluntary: Firm initiated
Z-0760-2019Class IITerminatedVoluntary: Firm initiated
Z-0761-2019Class IITerminatedVoluntary: Firm initiated