Presource Kit Total Joint Pack containing a double-wrapped basin component
Cardinal Health Inc. began a recall of Presource Kit Total Joint Pack containing a double-wrapped basin component. Product Usage: This kit is used on patients undergoing joint replacement knee or hip on . FDA classified it as Class II and gives the reason as "The double wrapping of this custom kit basin set is not supported by the current sterility assurance validation". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-0702-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Presource Kit Total Joint Pack containing a double-wrapped basin component. Product Usage: This kit is used on patients undergoing joint replacement knee or hip.
- Product code
- LRO General Surgery Tray
- Reason for recall
- The double wrapping of this custom kit basin set is not supported by the current sterility assurance validation.
- Root cause tag
- sterility
- FDA root cause
- Unknown/Undetermined by firm
- Action
- As only one customer is potentially impacted, in the interest of prompt action and notification, a teleconference was held on Friday, November 16, 2018 with the customer to communicate the impending actions. During this teleconference, the customer was informed that an HHE and field corrective action was in-process and that a study was being initiated to determine that their requested packaging configuration was acceptable. The customer also was informed that they should not continue to use the impacted double wrapped basin, and the components contained within, until the study was completed. Cardinal Health is following up on this teleconference with a written notification.
- Quantity in commerce
- 182 packs
- Distribution
- US state of Florida
- Product codes and lots
- Product code: SOP22TJCLK Lot numbers # 104405, 122793, and 987948
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0702-2019 | Class II | Terminated | Voluntary: Firm initiated |