Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0728-2019Class II

BiMobile UHMWPE Liner: ID-28mm/OD-50mm, Reference Number 184-260/02 The LINK BiMobile Dual Mobility System is...

Waldemar Link GmbH & Co. KG / initiated 2018-12-04 / Terminated / root cause: manufacturing process

Waldemar Link GmbH & Co. KG (Mfg Site) began a recall of BiMobile UHMWPE Liner: ID-28mm/OD-50mm, Reference Number 184-260/02 The LINK BiMobile Dual Mobility System is indicated for patients with mobility-limiting... on . FDA classified it as Class II and gives the reason as "The internal labeling included with these implants may include incorrect part numbers and could result in these incorrect labels being added to patient records". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0728-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Waldemar Link GmbH & Co. KG
Product
BiMobile UHMWPE Liner: ID-28mm/OD-50mm, Reference Number 184-260/02 The LINK BiMobile Dual Mobility System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The device is intended for cemented and cementless use.
Product code
MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Reason for recall
The internal labeling included with these implants may include incorrect part numbers and could result in these incorrect labels being added to patient records.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The firm, LinkBio Corp., Sent an "URGENT MEDICAL DEVICE RECALL" letter dated 12/4/2018 to its customers. The letter describes the product, problem and actions to be taken. The customers are instructed to do the following: Please identify your total inventory for this product. 1) Immediately cease use and segregate the above described product. 2) Complete the enclosed Consignee Reply Form, even if you have no inventory. 3) Return the above product and a copy of the Consignee Reply Form using Federal Express Account # 007183879 to: LinkBio Corporation 101 Roundhill Drive Rockaway, NJ 07446 Recall Return Authorization #: R-2018-02 Ensure that all users of the above products within your organization and other relevant persons have been notified of this safety information. If you have transferred the products to third parties, please pass on a copy of this information or notify the contact person indicated below. Should you have any questions or require assistance, please contact 1-973-625-1333 ext. 116 between 8:00 am - 5:30 pm (M-F) EST or email: [email protected] at any time.
Quantity in commerce
30
Distribution
US distribution to states of: AZ, FL, GA, LA, MI, MN, TX, and WA.
Product codes and lots
Lot# 1707121

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0728-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0728-2019Class IITerminatedVoluntary: Firm initiated
Z-0729-2019Class IITerminatedVoluntary: Firm initiated