PROXIMATE HCS Hemorrhoidal Circular Stapler
Ethicon Endo-Surgery Inc began a recall of PROXIMATE HCS Hemorrhoidal Circular Stapler Product Code: PPH01 PPH01: The PROXIMATE HCS Hemorrhoidal Circular Stapler and Accessories have application for... on . FDA classified it as Class II and gives the reason as "Hemorrhoidal Circular Stapler may have been assembled without a washer, this may potentially cause bleeding or soft tissue injury during use as the device may not fully cut when fired". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0695-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ethicon Endo-Surgery Inc
- Product
- PROXIMATE HCS Hemorrhoidal Circular Stapler Product Code: PPH01 PPH01: The PROXIMATE HCS Hemorrhoidal Circular Stapler and Accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculomucosal tissue.
- Product code
- GDW Staple, Implantable
- Reason for recall
- Hemorrhoidal Circular Stapler may have been assembled without a washer, this may potentially cause bleeding or soft tissue injury during use as the device may not fully cut when fired
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm, Ethicon, issued an "URGENT: MEDICAL DEVICE RECALL (REMOVAL)" to US consignees on 12/7/18, via UPS 2nd day mail. Ethicon Johnson & Johnson Affiliates in North America (Canada), Europe, LATAM, and Asia Pacific will notify their impacted consignees by providing the customer letter through post, email or sales representative visits. Letter identifies reason for recall, action to be taken: examine inventory and return recalled product for replacement; complete the Business Reply Form (BRF) confirming receipt of this notice and fax or email it to Stericycle at 1-844-524-3888 or [email protected] within three (3) business days. Please return the BRF even if you do not have the product lots subject to this recall.. Communicate the issue to relevant operating room or materials management personnel, or anyone else in the facility who needs to be informed. If any product subject to this action has been forwarded to another facility, contact that facility to arrange return. Customers are required to return all PROXIMATE PPH Hemorrhoidal Circular Stapler products subject to this recall that are in their inventory immediately. To receive replacement product, customers must return product subject to this recall by April 1, 2019. Any unused Proximate PPH Hemorrhoidal Circular Stapler product lots subject to this recall returned after April 1, 2019 will not be eligible for replacement. Contact your Ethicon sales representative or contact Ethicon Customer Support Center at 1-877-ETHICON (1-877-384-4266). The Customer Support Center is open Monday through Friday, 7:30 AM to 6:30 PM ET
- Quantity in commerce
- 480 units
- Distribution
- Worldwide Distribution: US (Nationwide) and to countries of: ARGENTINA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, ECUADOR, EGYPT, INDIA, INDONESIA, JAPAN, KOREA, KUWAIT, LEBANON MEXICO, NICARAGUA, PAKISTAN, SAUDI ARABIA, SINGAPORE, TAIWAN URUGUAY and VIETNAM.
- Product codes and lots
- Lot Codes: P9461W, P94K20
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0695-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0696-2019 | Class II | Terminated | Voluntary: Firm initiated |