Z-1125-2019Class II

cobas c 111 with ISE analyzers analyzer used in the following assays: a)Creatinine plus ver.2 Catalog Number...

Roche Diagnostics GmbH / initiated 2018-12-14 / Terminated / root cause: design

Roche Diagnostics Corporation began a recall of cobas c 111 with ISE analyzers analyzer used in the following assays: a)Creatinine plus ver.2 Catalog Number: 05401470190; b)Triglycerides Catalog Number... on . FDA classified it as Class II and gives the reason as "COBAS INTEGRA 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1125-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
cobas c 111 with ISE analyzers analyzer used in the following assays: a)Creatinine plus ver.2 Catalog Number: 05401470190; b)Triglycerides Catalog Number: 04657594190 - Product Usage: Tina-quant Apolipoprotein A-1 ver.2 A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis Tina-quant Albumin Gen.2 Immunoturbidimetric assay for the quantitative, in vitro determination of albumin in human urine, serum, plasma, and CSF on Roche/Hitachi cobas c systems
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
COBAS INTEGRA 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over
Root cause tag
design
FDA root cause
Device Design
Action
Roche Diagnostics issued on 12/14/18 "Urgent Medical Device Correction (UMDC) via UPS Ground (receipt required) Consignee Instructions: Manually add the EWCs listed in the Urgent Medical Device Correction (UMDC) on your COBAS INTEGRA 400 plus and cobas c 111 analyzers. If customers run the test combinations requiring needle assignment (See Table on page 3 of UMDC TP-00508), run those in batch mode until the updated TAS has been loaded on the COBAS INTEGRA 400 Plus analyzer. Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions.
Quantity in commerce
137 units
Distribution
US Nationwide distribution.
Product codes and lots
All

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1125-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1124-2019Class IITerminatedVoluntary: Firm initiated
Z-1125-2019Class IITerminatedVoluntary: Firm initiated