Z-0707-2019Class II

Siemens Biograph Horizon PET/CT System

Siemens Medical Solutions USA, Inc. / initiated 2018-11-30 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc. began a recall of Siemens Biograph Horizon PET/CT System, Product Usage: The Siemens Biograph systems are combined X-Ray Computed Tomography (CT) and Positron Emission... on . FDA classified it as Class II and gives the reason as "Improper design or specifications". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0707-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Biograph Horizon PET/CT System, Product Usage: The Siemens Biograph systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
Improper design or specifications.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens sent a letter to all their customers. The letter identified the affected product, problem and actions to be taken. Siemens Healthineers' service organization will contact you inQ1 of 2019 to schedule the replacement of the 70 amp CT circuit breaker, found within the PDCC cabinet of your Biograph Horizon system, with a 60 amp CT circuit breaker. Please ensure that this customer advisory notice is placed in the Biograph Horizon Operator's Manual for Examination and Acquisition and disseminated to all operators of Biograph Horizon. If this equipment is no longer in your possession, we kindly ask that you forward this letter to the new owner of the equipment and please inform Siemens Healthineers about the change in ownership. If you have any questions regarding this advisory notice, please contact your local Siemens Healthineers representative at the contact numbers provided below. " America: 1-800-888-7436 " Europe, Middle East, and Africa: +49 9131 940 4000 " Asia and Australia: +86 (21) 3811 2121
Quantity in commerce
171
Distribution
Worldwide Distribution - US Nationwide in the states of AZ, CA, CO, DE, FL, GA, ID, IL, IN, KY, LA, MD, ME, MN, NJ, NM, NV, NY, OH, OK, PA, TN, TX, and VA and the countries of Argentina, Australia, Bahrain, Belarus, Belgium, Brazil, Chile, China, Colombia, France, Germany, India, Iran, Ireland, Italy, Japan, Lebanon, Mexico, Netherlands, New Zealand, Philippines, South Africa, Spain, Sri Lanka, Taiwan, Turkey, and United Kingdom
Product codes and lots
Material Number: 10532746 - Serial #'s: 94401, 94405, 94406, 94408, 94409, 94413, 94414, 94415, 94420, 94421, 94422, 94423, 94424, 94425, 94426, 94427, 94429, 94430, 94431, 94432, 94433, 94434, 94435, 94436, 94438, 94440, 94441, 94442, 94443, 94444, 94445, 94447, 94448, 94451, 94452, 94457, 94459, 94460, 94461, 94462, 94463, 94464, 94466, 94467, 94473, 94476, 94477, 94480, 94481, 94482, 94483, 94484, 94485, 94487, 94491, 94492, 94493, 94495, 94496, 94497, 94498, 94499, 94500, 94502, 94504, 94505, 94506, 94508, 94510, 94512, 94513, 94516, 94517, 94518, 94519, 94520, 94521, 94522, 94525, 94530, 94531, 94532, 94533, 94534, 94537, 94538, 94540, 94541, 94542, 94543, 94547, 94549, 94550, 94551, 94555, 94556, 94558, 94559, 94560, 94562, 94563, 94565, 94566, 94567, 94569, 94574, 94576, 94580, and 94590 Material Number: 10532748 - Serial #'s: 94402, 94407, 94410, 94412, 94416, 94417, 94418, 94419, 94428, 94446, 94449, 94450, 94453, 94454, 94455, 94456, 94465, 94470, 94471, 94472, 94475, 94478, 94479, 94486, 94488, 94489, 94490, 94494, 94501, 94503, 94507, 94524, 94527, 94528, 94536, 94539, 94544, 94546, 94548, 94552, 94554, 94557, 94561, 94575, 94585, 94587, and 94593

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0707-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0707-2019Class IITerminatedVoluntary: Firm initiated