Z-0852-2019Class II

Stingray" Guidewire, labeled as 300cm. intended to facilitate placement of balloon dilatation catheters or...

Boston Scientific Corporation / initiated 2018-12-19 / Terminated / root cause: labeling and UDI

Boston Scientific Corporation began a recall of Stingray" Guidewire, labeled as 300cm. intended to facilitate placement of balloon dilatation catheters or other intravascular devices during percutaneous... on . FDA classified it as Class II and gives the reason as "A batch of 185cm Stingray Guidewires are incorrectly labeled as 300cm guidewires". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0852-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Stingray" Guidewire, labeled as 300cm. intended to facilitate placement of balloon dilatation catheters or other intravascular devices during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).
Product code
DQX Wire, Guide, Catheter
Reason for recall
A batch of 185cm Stingray Guidewires are incorrectly labeled as 300cm guidewires.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Boston Scientific notified customers on about 12/19/2018 via "Urgent Medical Device Removal - Immediate Action Required" notice. Instructions included to immediately discontinue use of and segregate affected product, complete and return the Reply Verification Tracking Form to receive a RGA to return affected product, return all affected product, notify customers of the recall if product was further distributed.
Quantity in commerce
49 units
Distribution
Nationwide distribution to AR, CT, FL, IA, IL, IN, KY, MA, MI, ND, NJ, NY, OH, OK, PA, SC, TX, VA, WI, WY.
Product codes and lots
Material Number (UPN) H749M3004A0, Batch Number 21558994, Expiration Date 11/30/2019

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0852-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0852-2019Class IITerminatedVoluntary: Firm initiated