Z-0789-2019Class II

Alinity i Processing Module, (a) Part Number 03R65-01 (b) Part Number A-30108552-01 (c) Part Number...

Abbott Medical / initiated 2018-12-20 / Terminated / root cause: manufacturing process

Abbott Gmbh & Co. KG began a recall of Alinity i Processing Module, (a) Part Number 03R65-01 (b) Part Number A-30108552-01 (c) Part Number A-30109694-01 Product Usage: The Alinity i processing... on . FDA classified it as Class II and gives the reason as "Abbott has identified an issue with the Alinity i Gear Pump Assembly (part number A 30108552 01) resulting in foaming/bubbling out of the bottle reservoir for Concentrated Wash Buffer and an unexpected amount of dried...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0789-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Alinity i Processing Module, (a) Part Number 03R65-01 (b) Part Number A-30108552-01 (c) Part Number A-30109694-01 Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Abbott has identified an issue with the Alinity i Gear Pump Assembly (part number A 30108552 01) resulting in foaming/bubbling out of the bottle reservoir for Concentrated Wash Buffer and an unexpected amount of dried residue of buffer. The foaming/bubbling is a result of an improper calibration of the pump s operating speed.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm initiated the recall by letter on 12/20/2018. The letter identified the affected product, problem and actions to be taken. The units will be serviced in the field. For questions contact Customer Service at 1-877-4ABBOTT. Customers outside the U.S., contact your local area Customer Service.
Quantity in commerce
434 units
Distribution
Worldwide Distribution
Product codes and lots
(a) Serial Numbers: Ai01001 - Ai01455, Ai01457, Ai01459 (b) all Serial Numbers (c) all Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0789-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0789-2019Class IITerminatedVoluntary: Firm initiated