NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi .0.8.521 9+PO1 Multi-Slice...

Neusoft Medical Systems Imp & EX / initiated 2019-01-09 / Terminated / root cause: software

NEUSOFT MEDICAL SYSTEMS IMP & EX began a recall of NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi .0.8.521 9+PO1 Multi-Slice CT Scanner System can be used as a whole body... on . FDA classified it as Class II and gives the reason as "For surview scan length more than 500mm,if the user aborts or skips the reconstructed surview, the image may have the wrong field of view (FOV). When user plans and performs subsequent clinical scan with the wrong FOV...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0773-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Neusoft Medical Systems Imp & EX
Product
NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi .0.8.521 9+PO1 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
For surview scan length more than 500mm,if the user aborts or skips the reconstructed surview, the image may have the wrong field of view (FOV). When user plans and performs subsequent clinical scan with the wrong FOV the actual start and end position may not be consistent with the planned positons.
Root cause tag
software
FDA root cause
Software design
Action
The firm, Neurosoft Medical Systems Co., Inc., issued a notice to its customers dated 11/26/2018. The notice described the product, problem and actions to be taken. The customers were instructed to do the following: Please plan surview length accurately according the clinical requirement, and not interrupt(abort or skip) scanning during the surview scan if the planning length is over 500mm. The defect will be remedied by updating software. Field Change Order related to software updated will be released to the affected systems in Jan, 2019. The affected systems will be upgraded with updated software in the field free of charge in 6 months after the FCO release. If you have any questions, contact Senior TQM Director at Tel: +86 024-2335831 1 or email: [email protected].
Quantity in commerce
3
Distribution
US Distribution in states of: KS, OK and MA.
Product codes and lots
Series number includes: N1281180242E, N1281170231E, N1281180248E

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0773-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0773-2019Class IITerminatedFDA Mandated