Z-0730-2019Class II
Sl Plus Mia Double Offset Adapter Right 60/25Mm, S&n 75004613
Smith & Nephew, Inc. / initiated 2019-01-07 / Terminated / root cause: manufacturing process
Smith & Nephew, Inc. began a recall of SL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613 on . FDA classified it as Class II and gives the reason as "The pin that holds the spring mechanism is not properly mounted. The pin could potentially disassemble". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0730-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Sl Plus Mia Double Offset Adapter Right 60/25Mm, S&n 75004613
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- The pin that holds the spring mechanism is not properly mounted. The pin could potentially disassemble.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm initiated the recall by email and letter on 01/07/2019. The firm requested the return of the product.
- Quantity in commerce
- 72 units
- Distribution
- NY, TN, Germany, Italy, Great Britain, Singapore
- Product codes and lots
- Lot Numbers: C63165, C63306, C63495
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0730-2019 | Class II | Terminated | Voluntary: Firm initiated |