Revolution CT systems with the SmartStep Option The system is intended for head, whole body, cardiac and...
GE Healthcare, LLC began a recall of Revolution CT systems with the SmartStep Option The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications on . FDA classified it as Class II and gives the reason as "On the Revolution CT systems equipped with the SmartStep Option, the Z location displayed on image annotation and on the gantry display may not be the actual location of the table and result in the system scanning at a...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0776-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Revolution CT systems with the SmartStep Option The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- On the Revolution CT systems equipped with the SmartStep Option, the Z location displayed on image annotation and on the gantry display may not be the actual location of the table and result in the system scanning at a location that is not what was confirmed by the operator
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- GE Healthcare sent a Notification letter dated October 18, 2018 to affected customers and sent a second letter dated October 30, 2018. The letters identified the affected product, problem and actions to be taken. The customers were instructed to do the following: ensure that all potential users in your facility are made aware of this notification and the recommended action, also stated: To avoid the issue, only use the 'In' button on the gantry controls and follow this required workflow: 1. Move the table back into the gantry using the 'In' button on the gantry controls . 2. Verify that the needle position is aligned using the internal laser alignment light. GE Healthcare will correct both forward production and impacted customer devices with a software update. For question contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity in commerce
- 131 units
- Distribution
- Worldwide Distribution - US (Nationwide Distribution) and countries of: Australia, Belgium, Canada, Estonia, Finland, France, Germany, Hong Kong, Ireland, Japan, Korea Republic of, Malaysia, Morocco, New Zealand, Norway, Saudi Arabia, Singapore, Slovakia, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, and United Kingdom.
- Product codes and lots
- SmartStep Option
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0776-2019 | Class II | Ongoing | FDA Mandated |