Z-0776-2019Class II

Revolution CT systems with the SmartStep Option The system is intended for head, whole body, cardiac and...

GE Healthcare / initiated 2018-07-12 / Open, Classified / root cause: software

GE Healthcare, LLC began a recall of Revolution CT systems with the SmartStep Option The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications on . FDA classified it as Class II and gives the reason as "On the Revolution CT systems equipped with the SmartStep Option, the Z location displayed on image annotation and on the gantry display may not be the actual location of the table and result in the system scanning at a...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0776-2019
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
Revolution CT systems with the SmartStep Option The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
On the Revolution CT systems equipped with the SmartStep Option, the Z location displayed on image annotation and on the gantry display may not be the actual location of the table and result in the system scanning at a location that is not what was confirmed by the operator
Root cause tag
software
FDA root cause
Software design
Action
GE Healthcare sent a Notification letter dated October 18, 2018 to affected customers and sent a second letter dated October 30, 2018. The letters identified the affected product, problem and actions to be taken. The customers were instructed to do the following: ensure that all potential users in your facility are made aware of this notification and the recommended action, also stated: To avoid the issue, only use the 'In' button on the gantry controls and follow this required workflow: 1. Move the table back into the gantry using the 'In' button on the gantry controls . 2. Verify that the needle position is aligned using the internal laser alignment light. GE Healthcare will correct both forward production and impacted customer devices with a software update. For question contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in commerce
131 units
Distribution
Worldwide Distribution - US (Nationwide Distribution) and countries of: Australia, Belgium, Canada, Estonia, Finland, France, Germany, Hong Kong, Ireland, Japan, Korea Republic of, Malaysia, Morocco, New Zealand, Norway, Saudi Arabia, Singapore, Slovakia, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, and United Kingdom.
Product codes and lots
SmartStep Option

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0776-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0776-2019Class IIOngoingFDA Mandated