Edward Lifesciences, Edwards Commander Delivery System: Model Number UDI Number 9600LDS20A (US)...
Edwards Lifesciences, LLC began a recall of Edward Lifesciences, Edwards Commander Delivery System: Model Number UDI Number 9600LDS20A (US) 00690103193930 9600LDS20J 00690103194852 9600LDS23A (US)... on . FDA classified it as Class II and gives the reason as "This action has been initiated to address complaint reports received for crimp balloon and inflation balloon component separation during TAVR procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0835-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Edwards Lifesciences LLC
- Product
- Edward Lifesciences, Edwards Commander Delivery System: Model Number UDI Number 9600LDS20A (US) 00690103193930 9600LDS20J 00690103194852 9600LDS23A (US) 00690103193947 9600LDS23J 00690103194869 9600LDS26A (US) 00690103193954 9600LDS26J 00690103194876 9600LDS29A (US) 00690103193961 9600LDS29J 00690103194883 9610TF20 00690103192995 9610TF23 00690103190458 9610TF26 00690103190465 9610TF29 00690103193008 Product Usage: The Edwards SAPIEN 3 THV, Edwards Commander delivery system and accessories are indicated for use in patients with severe, symptomatic, calcific aortic stenosis
- Product code
- NPT Aortic Valve, Prosthesis, Percutaneously Delivered
- Reason for recall
- This action has been initiated to address complaint reports received for crimp balloon and inflation balloon component separation during TAVR procedures.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- The firm began disseminating the notices by letter on 01/22/2019. The notices explained the issue and advised the consignee to follow the guidelines and instructions provided in the Edward SAPIEN3 Commander delivery system training material for device preparation and use. Updated instructions for use will be disseminated in the near future.
- Quantity in commerce
- 159,448 units
- Distribution
- WorldwideDistribution
- Product codes and lots
- (ALL LOTS Model Number UDI Number 9600LDS20A (US) 00690103193930 9600LDS20J 00690103194852 9600LDS23A (US) 00690103193947 9600LDS23J 00690103194869 9600LDS26A (US) 00690103193954 9600LDS26J 00690103194876 9600LDS29A (US) 00690103193961 9600LDS29J 00690103194883 9610TF20 00690103192995 9610TF23 00690103190458 9610TF26 00690103190465 9610TF29 00690103193008
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0835-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY