Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0826-2019Class II

21Y14 PushValve

Ottobock Orthopedic Industrie / initiated 2019-01-10 / Terminated / root cause: manufacturing process

Ottobock Orthopedic Industrie began a recall of 21Y14 PushValve Product Usage: The PushValves/ MagValves is intended exclusively for the prosthetic fitting of lower limb amputees. The valve is a threadless... on . FDA classified it as Class II and gives the reason as "Some of the installed valve-inserts have a diameter smaller than the specification". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0826-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ottobock Orthopedic Industrie
Product
21Y14 PushValve Product Usage: The PushValves/ MagValves is intended exclusively for the prosthetic fitting of lower limb amputees. The valve is a threadless one-way valve. It is used to evacuate air from the prosthetic socket.
Product code
ISY Joint, Knee, External Limb Component
Reason for recall
Some of the installed valve-inserts have a diameter smaller than the specification.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm initiated the recall by email on 01/10/2019 and followed with a letter on 01/11/2019. The letter identified the affected product, problem, actions to be taken and that a second letter with the replacement unit was forthcoming.
Quantity in commerce
538 units
Distribution
US Nationwide Distribution
Product codes and lots
The products with an article number listed below and delivered between May 1, 2017 and January 19, 2018

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0826-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0826-2019Class IITerminatedVoluntary: Firm initiated
Z-0827-2019Class IITerminatedVoluntary: Firm initiated
Z-0828-2019Class IITerminatedVoluntary: Firm initiated