daVinci Harmonic ACE Curved Shears
Intuitive Surgical, Inc. began a recall of daVinci Harmonic ACE Curved Shears on . FDA classified it as Class II and gives the reason as "During an internal inspection, Intuitive found that the sterile packaging of the da Vinci S/Si Harmonic ACE Inserts may become damaged during transit. The potential damage includes pinholes and /or small openings (less...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0841-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- daVinci Harmonic ACE Curved Shears
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- During an internal inspection, Intuitive found that the sterile packaging of the da Vinci S/Si Harmonic ACE Inserts may become damaged during transit. The potential damage includes pinholes and /or small openings (less than a few millimeters) on the pouch leading to a potential breach in sterility.
- Root cause tag
- packaging
- FDA root cause
- Under Investigation by firm
- Action
- A Stop Use Notification was sent to affected customers on December 19, 2018. This notification requested customers to return the device. The formal customer recall letter was sent on January 16, 2019.
- Quantity in commerce
- 43755 inserts
- Distribution
- US and Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Kuwait, Lebanon, Mexico, Monaco, Netherlands, Pakistan, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Korea, Spain, Taiwan, Turkey, United Kingdom
- Product codes and lots
- All Lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0841-2019 | Class II | Ongoing | Voluntary: Firm initiated |