Z-0841-2019Class II

daVinci Harmonic ACE Curved Shears

Intuitive Surgical, Inc. / initiated 2019-01-16 / Open, Classified / root cause: packaging

Intuitive Surgical, Inc. began a recall of daVinci Harmonic ACE Curved Shears on . FDA classified it as Class II and gives the reason as "During an internal inspection, Intuitive found that the sterile packaging of the da Vinci S/Si Harmonic ACE Inserts may become damaged during transit. The potential damage includes pinholes and /or small openings (less...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0841-2019
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Intuitive Surgical, Inc.
Product
daVinci Harmonic ACE Curved Shears
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
During an internal inspection, Intuitive found that the sterile packaging of the da Vinci S/Si Harmonic ACE Inserts may become damaged during transit. The potential damage includes pinholes and /or small openings (less than a few millimeters) on the pouch leading to a potential breach in sterility.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
A Stop Use Notification was sent to affected customers on December 19, 2018. This notification requested customers to return the device. The formal customer recall letter was sent on January 16, 2019.
Quantity in commerce
43755 inserts
Distribution
US and Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Kuwait, Lebanon, Mexico, Monaco, Netherlands, Pakistan, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Korea, Spain, Taiwan, Turkey, United Kingdom
Product codes and lots
All Lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0841-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0841-2019Class IIOngoingVoluntary: Firm initiated