Z-0961-2019Class II

Alcon or Endure Ophthalmic Microscope

Alcon Laboratories / initiated 2018-07-17 / Terminated / root cause: design

Alcon Research Laboratories Ltd. began a recall of Alcon or Endure Ophthalmic Microscope Product Usage: Used for low magnification visualization during ophthalmic surgical procedures for cataract, retina and... on . FDA classified it as Class II and gives the reason as "Potential for the optical head to detach from the stand due to the lack of thread adhesive". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0961-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alcon Laboratories
Product
Alcon or Endure Ophthalmic Microscope Product Usage: Used for low magnification visualization during ophthalmic surgical procedures for cataract, retina and cornea.
Product code
HRM Microscope, Operating & Accessories, Ac-Powered, Ophthalmic
Reason for recall
Potential for the optical head to detach from the stand due to the lack of thread adhesive.
Root cause tag
design
FDA root cause
Device Design
Action
On July 17, 2018 the firm sent letters to their consignees with the following instructions: To assist us in this Voluntary Medical Device Correction, please take the following steps: 1. Inspect your microscope prior to continued use per the instruction above. 2. If either the Functional or Visual tests have failed, stop use of the device immediately and notify Alcon at 1-800-832-7827. 3. Forward this information immediately to all departments within your organization who may be using the Alcon or Endure LuxOR E71 Ophthalmic Microscope and Alcon or Endure LuxOR E71 Q-Vue Ophthalmic Microscope. 4. If applicable, provide a copy of this notification to any other organization I location to which the Alcon or Endure LuxOR E71 Ophthalmic Microscope and Alcon or Endure LuxOR E71 Q-Vue Ophthalmic Microscope has been sold or moved. 5. Return the attached Response Form via fax or email to Alcon.
Quantity in commerce
544 (399 units to USA)
Distribution
Worldwide Distribution - US Nationwide Distribution in the states of AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, Ml, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, Puerto Rico and US Virgins Islands Australia, Austria, Bahamas, Belarus, Canada, Chile, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador, El Salvador, Finland, Germany, Guatemala, India, Indonesia, Italy, Jamaica, Japan, Kenya, Lebanon, Mexico, Netherland Antilles, Norway, Peru, Philippines, Poland, Romania, Slovakia, Spain, Sri Lanka, Sweden, Taiwan, Thailand, Ukraine, United Kingdom, and Vietnam
Product codes and lots
Model: LuxOr E71 and E71 Q-Vue 8065752304, 8065752323, 8065752332, and 8065752333 US Serial numbers: 22013, 1401148103X, 1401148201X, 1401148204X, 1401554606X, 1401845004X, 1401845005X, 1401974102X, 1401974104X, 1401974105X, 1401977002X, 1401977003X, 1402067603X, 1402067604X, 1402067605X, 1402149501X, 1402149503X, 1402149504X, 1402167704X, 1402249701X, 1402249702X, 1402249703X, 1402249803X, 1402314601X, 1402314701X, 1402314705X, 1402355101X, 1402677604X, 1402804905X, 1402877605X, 1402877701X, 1402877702X, 1402877801X, 1402877804X, 1402959102X, 1402959201X, 1402959202X, 1402959204X, 1403031005X, 1403133802X, 1403133804X, 1403133805X, 1403353401X, 1403353402X, 1403353403X, 1403353404X, 1403353405X, 1403571101X, 1403571102X, 1403571103X, 1403571104X, 1403571105X, 1403701202X, 1403701203X, 1403701204X, 1403701205X, 1403701206X, 19941, 20214, 20349, 20381, 20585, 20652, 20702, 20714, 20722, 20781, 20832, 20862, 20878, 20878-2, 20879, 20933, 21017, 21019, 21032, 21104, 21171, 21173, 21217, 21270, 21463, 21516, 21567, 21590, 21630, 21639, 21654, 21678, 21749, 21750, 21815, 21849, 21867, 21940, 21972, 22003, 22026, 22029, 22088, 22088-2, 22134, 22182, 22231, 2713, 30131, 30216, 30363, 30363-2, 30766, 31172, 31172-2, 3157, 31646, 31745, 31780, 31873, 31873-2, 31873-3, 32004, 32214, 32347, 32384, 32384-2, 32400 32420, 32420-2, 32439, 32453, 32553, 32568, 32643, 32694, 32694-2, 32698, 32743, 32748, 32755, 32771, 32785, 32785-2, 32785-3, 32875, 32886, 32886-2, 32896, 32909, 32923, 32923-2, 32994, 33028, 33039, 33176, 33176-2, 33263, 33270, 33313, 33325, 33348, 33348-2, 33355, 33358, 33392, 33392-2, 33440, 33442, 33454, 33460, 33464, 33498, 33498-2, 33522, 33535, 33592, 33598, 33606, 33606-2, 33606-3, 33606-4, 33653, 33769, 33775, 33783, 33849, 33850, 33850-2, 33892, 33904, 33926, 33966, 33999,  34030, 34030-2, 34033, 34099, 34099-2, 34173, 34203, 34204, 34280, 3431 1, 34312, 34352, 34352-2, 34371, 34458, 34458-2, 34468, 34556, 34562, 34564, 34597, 34616, 34627, 34627-2, 34627-3, 34679, 34680, 34682, 34682-2, 34686A-2, 34699, 34700, 34701, 34780, 34884, 34911, 34951, 34956, 34974, 34999, 34999-2, 35010, 35039, 35061, 35063, 35063-2, 35064, 35120, 35144, 35149, 35162, 35163, 35203, 35266, 35288, 35289, 35316, 35330, 35334, 5382, 35382-2, 35383, 35433, 35433-2, 35433-3, 35433-4, 35450, 35462, 35511, 35553, 35572, 35602, 35615, 35616, 35616-2, 35644, 35690, 35690-2, 35696, 35742, 35742-2, 35785, 35786, 35789, 35789-2, 35791, 35827, 35833, 35855, 35860, 35860-2, 35886, 35893-4, 35919, 35967, 35969, 35969-2, 35981, 35985, 35985-2, 35993, 36013, 36013-2, 36014, 36021, 36022, 36031, 36031-2, 36077, 36078, 40220 :002, 40220:004, 40220:006, 40232:00, 40476, 40514:001, 40516 :001, 40518:001, 40545 :003, 40547:003, 40548:003, 40609:003, 40612:003, 40680:001, 40681 :001, 40684:001, 40685 :001, 40687:001, 40688:001, 40769:001, 40770:001, 40773 :001, 40776:001, 40778 :001, 40781:001, 40783:001, 40797 :001, 40798 :001, 40801 :001, 40803:001, 40804:001, 40806 :001, 40807:001, 40837:003, 41147 :001,  41151 :001, 41153:001, 41190:001, 41212:001, 41256 :001, 4157 :001, 41266:001, 41270 :001, 41272 :001, 41273:001, 41481 :001, 41612:001, 41613 :001, 41842:001, 41853 :001, 41858:001, 41859 :001, 41860 :001, 41865 :001, 42109 :001, 42111 :001, 42321:001, 42334:001, 42345 :001, 42346 :001, 42497 :001, 42498 :001, 42500:001, 42507 :001, 42883 :001, 43178 :001, 43179 :001, 43241 :001, 43638 :001, 3640 :001, 43680 :001, 43685 :001, 43690 :001, 43691 :001, 3692 :001, 43693 :001, 43813 :001, 43814:001, 43816 :001, 43821 :001, 44025 :001, 44063 :001, 44083 :001, 44084 :001, 4090 :001, 44097 :001, 44299 :001, 44306 :001, 44308 :001, 44310 :001, 44313 :001, 44314 :001, 44357 :001, 44358 :001, A005, A006, A012, A014, A016 and A018 OUS Serial Numbers: 41242:001, 41275 :001, 41848 :001, 1401148001X, 1401148101X, 1401148104X, 1402643103X, 1402314404X, 1402314405X, 1403918102X, 1502158505X, 42119:001, 42994 :001, 44040:001, 43232 :001, 1401975304X, 1402388004X, 43238:001, 43240:001, 41841 :001, 41258 :001, 1401554610X, 1401954603X, 1402067602X, 1402677602X, 41891:001, 43000 :001, 43246:001, 43487 :001, 43546 :001, 43547 :001, 43566 :001, 1401148203X, 1402250002X, 1403600301X, 1403600305X, 1501088505X, 1501247401X, 1501247403X, 1401148105X, 43947 :001, 44027 :001, 41225 :001, 42348 :001, 42998 :001, 43002 :001, 43167 :001, 1401148202X, 41140 :001, 42201:001, 43045 :001, 43400 :001, 42793 :001, 42796 :001, 1501355802X, 43548 :001, 1401554603X, 1401579101X, 1401579103X, 1401831601X, 1401991002X, 1403211805X, 1502004802X, 43557:001, 43633:001, 43645:001, 44058:001, 44060:001, 44093:001, 44095:001, 44096:001, 1402643102X, 42115:001, 1401975302X, 1402677603X, 1402677605X, 1403133905X, 1501088502X, 1501088506X, 1501088507X, 1501088508X, 1501247402X, 40610 :003, 41070:001, 41073:001, 41143:001, 41226:001, 41240 :001, 41627 :001, 41628:001, 43234:001, 43236 :001, 43237 :001, 43239:001, 44054:001, 1402388002X, 101200157, 1403331702X, 1403331805X, 43256 :001,  43549 :001, 41619 :001, 42797 :001, 43172 :001, 43301 :001, 41631:001, 41632 :001, 1402677601X, 41616 :001, 1402249804X, 1402314402X, 40611 :003, 40853:002, 41261:001, 41854:001, 42114:001, 42347 :001, 42588:001, 42608 :001, 43486 :001, 43644 :001, 1401554608X, 41688:001, 43173 :001, 43177-001, 43229:001, 43683:001, 42105 :001, 42107:001, 42108 :001, 43789:001, 43790 :001, 43791 :001, 44048 :001, 44049 :001, 44050 :001, 42503 :001, 1401554509X, 40805 :001, 41482 :001, 41611 :001, 41846 :001, and 42110 :001.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0961-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0961-2019Class IITerminatedVoluntary: Firm initiated