Z-0962-2020Class II
PTFE Wire Guide, Guidewire, Global Product Number 638813 635413-10
Cook Incorporated / initiated 2019-01-04 / Terminated / root cause: manufacturing process
Cook Inc. began a recall of PTFE Wire Guide, Guidewire, Global Product Number 638813 635413-10 on . FDA classified it as Class II and gives the reason as "The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0962-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- PTFE Wire Guide, Guidewire, Global Product Number 638813 635413-10
- Product code
- OCY Endoscopic Guidewire, Gastroenterology-Urology
- Reason for recall
- The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
- Quantity in commerce
- 3929
- Distribution
- Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
- Product codes and lots
- 7832498 7832499 7832500 7925385 7925387 7930517 8090868 8287793 8287794 8287799 8400121 8548813 8548815 NS8548814 7760235 7881599 8062433 8074027 8290383 8290384 8404559 8404562 7999663 8077582 8077583 8065959 8073949 8074014 8074015
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0959-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0960-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0961-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0962-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0963-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0964-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0965-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0966-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0967-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0968-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0969-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0970-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0971-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0972-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0973-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0974-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0975-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0976-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0977-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0978-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.