AMSCO 3000 Series Washer/Disinfector Model # 3052
Steris Corporation began a recall of AMSCO 3000 Series Washer/Disinfector Model # 3052 on . FDA classified it as Class II and gives the reason as "The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0995-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STERIS Corporation
- Product
- AMSCO 3000 Series Washer/Disinfector Model # 3052
- Product code
- MEC Disinfector, Medical Devices
- Reason for recall
- The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection.
- Root cause tag
- software
- FDA root cause
- Software Manufacturing/Software Deployment
- Action
- Urgent Device Field Correction notification letters dated 12/20/18 were sent to customers.
- Quantity in commerce
- 11
- Distribution
- The products were distributed to the following US states: AL, AR, CA, CO, FL, GA, IL, LA, MA, MD, ME, MI, NC, NJ, NM, NY, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The products were distributed to the following foreign countries: Bahrain, Brazil, Canada, China, Saudi Arabia, South Korea, and Taiwan.
- Product codes and lots
- Serial Numbers 3602014006 3631618006 3628818009 3629518008 3633018010 3626718010 3630918008 3630918010 3617717013 3620016019 3623516006
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0995-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0996-2019 | Class II | Terminated | Voluntary: Firm initiated |