Z-0996-2019Class II

AMSCO 5000 Series Washer/Disinfector Model # 5052

STERIS Corporation / initiated 2018-12-20 / Terminated / root cause: software

Steris Corporation began a recall of AMSCO 5000 Series Washer/Disinfector Model # 5052 on . FDA classified it as Class II and gives the reason as "The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0996-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
AMSCO 5000 Series Washer/Disinfector Model # 5052
Product code
MEC Disinfector, Medical Devices
Reason for recall
The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
Urgent Device Field Correction notification letters dated 12/20/18 were sent to customers.
Quantity in commerce
68
Distribution
The products were distributed to the following US states: AL, AR, CA, CO, FL, GA, IL, LA, MA, MD, ME, MI, NC, NJ, NM, NY, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The products were distributed to the following foreign countries: Bahrain, Brazil, Canada, China, Saudi Arabia, South Korea, and Taiwan.
Product codes and lots
Serial Numbers 3627418005 3601518006 3612718013 3633817009 3632318011 3626718014 3625114008 3600515028 3600515029 3600515030 3628516022 3626718013 3631618008 3627418007 3624514006 3628818001 3608518007 3634517006 3630218008 3605316016 3600917004 3628218007 3628818003 3629518009 3626718009 3626718011 3627418001 3627418002 3630218012 3630218013 3630918013 3632318012 3626718016 3634517005 3631414004 3603618005 3607217008 3630318013 3626817008 3628317012 3616615026 3630918014 3633018006 3628218009 3632318009 3628218011 3623615011A 3601518012 3612817007 3627418009 3628218006 3627418008 3631618007 3617315011 3614818009 3624017008 3629215012 3614516007 3628218010 3610216025 3634313010 3611715016 3628818005 3615617008 3624017009 3616516020 3627418004 3627418006

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0996-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0995-2019Class IITerminatedVoluntary: Firm initiated
Z-0996-2019Class IITerminatedVoluntary: Firm initiated