Roadrunner Hydrophilic PC Wire Guide, Guidewire, Global Product Number G18629, G17866
Cook Inc. began a recall of Roadrunner Hydrophilic PC Wire Guide, Guidewire, Global Product Number G18629, G17866 on . FDA classified it as Class II and gives the reason as "The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0963-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Roadrunner Hydrophilic PC Wire Guide, Guidewire, Global Product Number G18629, G17866
- Product code
- OCY Endoscopic Guidewire, Gastroenterology-Urology
- Reason for recall
- The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
- Quantity in commerce
- 13388 total
- Distribution
- Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
- Product codes and lots
- 7853231 7853232 7853233 7853239 7853240 7853241 7853242 7853243 7853245 7853247 7901140 7901141 7982113 7982118 7982133 8050291 8241613 8241616 8252179 8252181 8252186 8258553 8283322 8407934 8460985 8474895 8474898 8474905 8474906 8474908 8474909 8474910 7936207
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0959-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0960-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0961-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0962-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0963-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0964-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0965-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0966-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0967-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0968-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0969-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0970-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0971-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0972-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0973-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0974-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0975-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0976-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0977-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0978-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.