Angled Tip Ureteral Catheter Set with Bentson PTFE Wire Guide, Ureteral Catheter, Global Product No. G19108...

Cook Incorporated / initiated 2019-01-04 / Terminated / root cause: manufacturing process

Cook Inc. began a recall of Angled Tip Ureteral Catheter Set with Bentson PTFE Wire Guide, Ureteral Catheter, Global Product No. G19108 G14587 G15302 G14326 on . FDA classified it as Class II and gives the reason as "The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0966-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Angled Tip Ureteral Catheter Set with Bentson PTFE Wire Guide, Ureteral Catheter, Global Product No. G19108 G14587 G15302 G14326
Product code
OCY Endoscopic Guidewire, Gastroenterology-Urology
Reason for recall
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
Quantity in commerce
13388 total
Distribution
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Product codes and lots
8018728 8080823 8314382 8409523 8023194 7857489

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0966-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0959-2020Class IITerminatedVoluntary: Firm initiated
Z-0960-2020Class IITerminatedVoluntary: Firm initiated
Z-0961-2020Class IITerminatedVoluntary: Firm initiated
Z-0962-2020Class IITerminatedVoluntary: Firm initiated
Z-0963-2020Class IITerminatedVoluntary: Firm initiated
Z-0964-2020Class IITerminatedVoluntary: Firm initiated
Z-0965-2020Class IITerminatedVoluntary: Firm initiated
Z-0966-2020Class IITerminatedVoluntary: Firm initiated
Z-0967-2020Class IITerminatedVoluntary: Firm initiated
Z-0968-2020Class IITerminatedVoluntary: Firm initiated
Z-0969-2020Class IITerminatedVoluntary: Firm initiated
Z-0970-2020Class IITerminatedVoluntary: Firm initiated
Z-0971-2020Class IITerminatedVoluntary: Firm initiated
Z-0972-2020Class IITerminatedVoluntary: Firm initiated
Z-0973-2020Class IITerminatedVoluntary: Firm initiated
Z-0974-2020Class IITerminatedVoluntary: Firm initiated
Z-0975-2020Class IITerminatedVoluntary: Firm initiated
Z-0976-2020Class IITerminatedVoluntary: Firm initiated
Z-0977-2020Class IITerminatedVoluntary: Firm initiated
Z-0978-2020Class IITerminatedVoluntary: Firm initiated