Z-0971-2020Class II

Urethral Dilator Set, Global Product No. G14185

Cook Incorporated / initiated 2019-01-04 / Terminated / root cause: manufacturing process

Cook Inc. began a recall of Urethral Dilator Set, Global Product No. G14185 on . FDA classified it as Class II and gives the reason as "The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0971-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Urethral Dilator Set, Global Product No. G14185
Product code
KOE Dilator, Urethral
Reason for recall
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
Quantity in commerce
13388 total
Distribution
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Product codes and lots
8401849 8404945 8404946 8417702 8404946X

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0971-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0959-2020Class IITerminatedVoluntary: Firm initiated
Z-0960-2020Class IITerminatedVoluntary: Firm initiated
Z-0961-2020Class IITerminatedVoluntary: Firm initiated
Z-0962-2020Class IITerminatedVoluntary: Firm initiated
Z-0963-2020Class IITerminatedVoluntary: Firm initiated
Z-0964-2020Class IITerminatedVoluntary: Firm initiated
Z-0965-2020Class IITerminatedVoluntary: Firm initiated
Z-0966-2020Class IITerminatedVoluntary: Firm initiated
Z-0967-2020Class IITerminatedVoluntary: Firm initiated
Z-0968-2020Class IITerminatedVoluntary: Firm initiated
Z-0969-2020Class IITerminatedVoluntary: Firm initiated
Z-0970-2020Class IITerminatedVoluntary: Firm initiated
Z-0971-2020Class IITerminatedVoluntary: Firm initiated
Z-0972-2020Class IITerminatedVoluntary: Firm initiated
Z-0973-2020Class IITerminatedVoluntary: Firm initiated
Z-0974-2020Class IITerminatedVoluntary: Firm initiated
Z-0975-2020Class IITerminatedVoluntary: Firm initiated
Z-0976-2020Class IITerminatedVoluntary: Firm initiated
Z-0977-2020Class IITerminatedVoluntary: Firm initiated
Z-0978-2020Class IITerminatedVoluntary: Firm initiated