Z-0972-2020Class II

Bander Ureteral Diversion Open-End Stent Set, Global Product No. G14822 G14823

Cook Incorporated / initiated 2019-01-04 / Terminated / root cause: manufacturing process

Cook Inc. began a recall of Bander Ureteral Diversion Open-End Stent Set, Global Product No. G14822 G14823 on . FDA classified it as Class II and gives the reason as "The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0972-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Bander Ureteral Diversion Open-End Stent Set, Global Product No. G14822 G14823
Product code
GBL Catheter, Ureteral, General & Plastic Surgery
Reason for recall
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
Quantity in commerce
13388 total
Distribution
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Product codes and lots
8011030 8033568 8063430 8063431 8150491 8417631 8417632 8587162 8587163 8590532 8590533 7990830 8003084 8033571 8063428 8590520 8590522 8629475 NS8629473

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0972-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0959-2020Class IITerminatedVoluntary: Firm initiated
Z-0960-2020Class IITerminatedVoluntary: Firm initiated
Z-0961-2020Class IITerminatedVoluntary: Firm initiated
Z-0962-2020Class IITerminatedVoluntary: Firm initiated
Z-0963-2020Class IITerminatedVoluntary: Firm initiated
Z-0964-2020Class IITerminatedVoluntary: Firm initiated
Z-0965-2020Class IITerminatedVoluntary: Firm initiated
Z-0966-2020Class IITerminatedVoluntary: Firm initiated
Z-0967-2020Class IITerminatedVoluntary: Firm initiated
Z-0968-2020Class IITerminatedVoluntary: Firm initiated
Z-0969-2020Class IITerminatedVoluntary: Firm initiated
Z-0970-2020Class IITerminatedVoluntary: Firm initiated
Z-0971-2020Class IITerminatedVoluntary: Firm initiated
Z-0972-2020Class IITerminatedVoluntary: Firm initiated
Z-0973-2020Class IITerminatedVoluntary: Firm initiated
Z-0974-2020Class IITerminatedVoluntary: Firm initiated
Z-0975-2020Class IITerminatedVoluntary: Firm initiated
Z-0976-2020Class IITerminatedVoluntary: Firm initiated
Z-0977-2020Class IITerminatedVoluntary: Firm initiated
Z-0978-2020Class IITerminatedVoluntary: Firm initiated