Kwart Retro-Inject Ureteral Stent Set, Global Product No. G15016 G14836 G14885 G14887 G17151
Cook Inc. began a recall of Kwart Retro-Inject Ureteral Stent Set, Global Product No. G15016 G14836 G14885 G14887 G17151 on . FDA classified it as Class II and gives the reason as "The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0974-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Kwart Retro-Inject Ureteral Stent Set, Global Product No. G15016 G14836 G14885 G14887 G17151
- Product code
- FAD Stent, Ureteral
- Reason for recall
- The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
- Quantity in commerce
- 13388 total
- Distribution
- Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
- Product codes and lots
- 8011027 8183322 8037660 8030233 8018054 7994816
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0959-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0960-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0961-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0962-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0963-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0964-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0965-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0966-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0967-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0968-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0969-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0970-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0971-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0972-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0973-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0974-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0975-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0976-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0977-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0978-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.