Sof-Flex Double Pigtail Ureteral Stent Set, Global Product No. G15076 G14906 G14840 G14865 G17128
Cook Inc. began a recall of Sof-Flex Double Pigtail Ureteral Stent Set, Global Product No. G15076 G14906 G14840 G14865 G17128 on . FDA classified it as Class II and gives the reason as "The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0975-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Sof-Flex Double Pigtail Ureteral Stent Set, Global Product No. G15076 G14906 G14840 G14865 G17128
- Product code
- FAD Stent, Ureteral
- Reason for recall
- The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
- Quantity in commerce
- 13388 total
- Distribution
- Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
- Product codes and lots
- NS7949093 NS8077953 NS8077954 NS8077955 8070691 8183342 8183343 8070693 8172134 8187189 NS8070692 8424304 7990841
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0959-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0960-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0961-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0962-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0963-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0964-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0965-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0966-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0967-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0968-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0969-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0970-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0971-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0972-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0973-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0974-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0975-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0976-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0977-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0978-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.