Z-1037-2019Class II
Nelaton Urinary Catheter, disposable, sterile, 100 eaches per sales unit. Labeled as: CH16/40CM and CH14/40CM
ConvaTec Limited / initiated 2019-01-09 / Open, Classified / root cause: packaging
ConvaTec, Inc began a recall of Nelaton Urinary Catheter, disposable, sterile, 100 eaches per sales unit. Labeled as: CH16/40CM and CH14/40CM Product Usage: Urinary Catheters are sterile... on . FDA classified it as Class II and gives the reason as "An internal assessment of packaging confirmed the potential for a breach in the sterile barrier". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1037-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- ConvaTec Limited
- Product
- Nelaton Urinary Catheter, disposable, sterile, 100 eaches per sales unit. Labeled as: CH16/40CM and CH14/40CM Product Usage: Urinary Catheters are sterile disposable devices intended to provide an intermittent pathway for drainage of fluids from the bladder. They are intended to be inserted intermittently through the urethra and up to the bladder enabling urine to drain. They are designed for short-term, intermittent transient use only and are not intended to be left in situ (there is no balloon on the tip to allow the catheter to be retained within the bladder). They are intended for single use only. They are intended to be removed once the bladder is fully drained. The goal of intermittent catheterization using Urology Catheters is to completely empty the bladder and prevent Urinary Tract Infections (UTIs).
- Product code
- KOD Catheter, Urological
- Reason for recall
- An internal assessment of packaging confirmed the potential for a breach in the sterile barrier.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- ConvaTec notified customers on about 01/09/2019 via "URGENT: VOLUNTARY MEDICAL DEVICE RECALL" letter. Instructions included how to identify affected product on the primary packaging and/or shipping carton as well as specific instructions for distributors, retailers, and end users (hospitals, EMT services, others). Distributors were instructed to inspect their inventory for affected products, perform a count of affected product in inventory, complete and return the response form, contact Customer Service at 1-800-582-6514 or [email protected] to report affected inventory on hand, forward the recall notice package to customers if product was further distributed, and provide a complete list of all consignees to allow ConvaTec to perform effectiveness checks. Retailers were instructed to immediately stop distributing and quarantine any affected inventory, perform a count of affected product in inventory, complete and return the response form, contact Customer Service at 1-800-582-6514 or [email protected] to report affected inventory on hand, and post page one of the recall notice in a conspicuous location in the store. End users (hospitals, EMT service, others) were instructed to inspect inventory to confirm if any affected product codes are in inventory, perform a count of affected product in inventory, complete and return the response form, contact Customer Service at 1-800-422-8811 or [email protected] to report affected inventory on hand.
- Quantity in commerce
- 5,744,525 units total
- Distribution
- Worldwide Distribution - US Nationwide in the states of IL, MI, PA, WI. and the countries of Algeria, Australia, Austria, Belgium, Canada, Croatia, Czechia, Czech Republic, Denmark, Egypt, Estonia, Finland, Faroe Islands, France, Germany, Greece, Hong Kong, Hungary, Iceland, Israel, Italy, Kenya, Korea, Kuwait, Lebanon, Myanmar, Netherlands, New Zealand, Norway, Poland, Qatar, Singapore, Spain, Slovakia, South Africa, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, United States, Vietnam.
- Product codes and lots
- Product Codes/REF Numbers: 505599, 505603; Lot Codes: 222348, 222349
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1035-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1036-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1037-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1038-2019 | Class II | Ongoing | Voluntary: Firm initiated |