Z-0174-2020Class II
Semi-Rigid Suction Canister, 3000cc. Model Numbers OR530 and OR530H. For OR530, the product comes in a case...
MEDLINE INDUSTRIES, LP - Northfield / initiated 2019-01-31 / Open, Classified / root cause: manufacturing process
Medline Industries Inc began a recall of Semi-Rigid Suction Canister, 3000cc. Model Numbers OR530 and OR530H. For OR530, the product comes in a case with two boxes. One box includes 50 eaches of... on . FDA classified it as Class II and gives the reason as "Potential for the Semi-Rigid canister lid to fragment during use". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0174-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Semi-Rigid Suction Canister, 3000cc. Model Numbers OR530 and OR530H. For OR530, the product comes in a case with two boxes. One box includes 50 eaches of liners in a poly bag and the second box includes 50 eaches of lids in a poly bag. For OR530H, the customer receives the number of liners and lids purchased in one case without the individual boxes. The "H" denotes a single quantity when being ordered.
- Product code
- KDQ Bottle, Collection, Vacuum
- Reason for recall
- Potential for the Semi-Rigid canister lid to fragment during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On about 01/31/2018 Medline notified customers via "URGENT RECALL" letter. Instructions included to immediately check inventory for and quarantine all affected product, complete and return the response form, and notify customers if the product has been further distributed and instruct them to return products to Medline Industries, Inc. Once the response form is received, return labels will be provided to return all affected product. For questions call 866-359-1704.
- Quantity in commerce
- 750 eaches
- Distribution
- Nationwide distribution to AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MO, MT, NC, NH, NJ, NV, NY, OH, PA, RI, SC, TN, TX, VA, VT, WI.
- Product codes and lots
- Model Numbers OR530, OR530H; Lot Numbers 20180914-1, 20180913-1, 20180916-1, 20180929-1