Monaco Radiation Treatment Planning (RTP) System
Elekta, Inc. began a recall of Monaco Radiation Treatment Planning (RTP) System on . FDA classified it as Class II and gives the reason as "If Improve Target Dose was chosen as an optimization model in a previous treatment session, Monaco will automatically use this optimization model again when proceeding with the online plan adaptation of a completion...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0965-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Monaco Radiation Treatment Planning (RTP) System
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- If Improve Target Dose was chosen as an optimization model in a previous treatment session, Monaco will automatically use this optimization model again when proceeding with the online plan adaptation of a completion plan when it should not.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Elekta notified customers on about 2/08/2019 via "URGENT IMPORTANT FIELD SAFETY NOTIFICATION." Instructions included to ensure the optimization model is set to "Reproduce Goal Dose" when creating a completion plan. Customers were also requested to post the recall notice in an accessible place for all users to see, advise the appropriate personnel working with the device of the recall, and to complete and return the Acknowledgement Form. The issue will be resolved in Monaco release 5.40.01.
- Quantity in commerce
- 14 devices
- Distribution
- Domestic distribution to Wisconsin, Texas. International distribution to Denmark, England, Germany, Netherlands, Sweden.
- Product codes and lots
- Software version 5.40
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0965-2019 | Class II | Terminated | Voluntary: Firm initiated |