Z-0867-2019Class II

Vitality Spinal Fixation System. Spinal Interlaminal, Appliance, Fixation, (non-sterile). Pedicle screw...

Highridge Medical, LLC / initiated 2018-07-30 / Terminated / root cause: manufacturing process

Zimmer Biomet Spine Inc. began a recall of Vitality Spinal Fixation System. Spinal Interlaminal, Appliance, Fixation, (non-sterile). Pedicle screw fixation is indicated for skeletally mature patients... on . FDA classified it as Class II and gives the reason as "The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0867-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Highridge Medical, LLC
Product
Vitality Spinal Fixation System. Spinal Interlaminal, Appliance, Fixation, (non-sterile). Pedicle screw fixation is indicated for skeletally mature patients and for adolescent patients. For use in Orthopedic medical specialty area. Product Usage: The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue. These devices are indicated as an adjunct to fusion for various indications.
Product code
KWP Appliance, Fixation, Spinal Interlaminal
Reason for recall
The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for various indications
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On August 30, 2018 a Urgent Medical Recall notification letter was sent to Distributors and Hospital Risk Managers notifying them of the Zimmer Biomet Vitality Shear-off Set Screw recall. The notification letter described the potential risks associated to the recall and steps they needed to follow to mitigate those potential risk. The responisbilities as described by Zimmer Biomet to the customers is as follows: Your Responsibilities 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to [email protected] within three (3) days. b. For each return, send a copy of Attachment 1 to [email protected]. c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. d. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to [email protected]. a. Identify whether there are any additional hospitals that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. b. If there are no additional accounts to notify, please indicate that there are no additional accounts, or indicate None or NA on the form. 5. Retain a copy of your Inventory Return Certification and product return forms for your records in the event of a compliance audit of your facility. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-447-3625 between 8:00 am and 5:00pm MST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an o
Quantity in commerce
16,429 devices (Thoracolumbosacral pedicle screw system,Spinal Interlaminal & Spinal Intervertebral Body, both appliances for fixation)
Distribution
Distributed Nationwide: AL, CA, CO, FL, GA, HI, IA, ID, IL, IN, LA, MA, MD, MI, MN, NC, NJ, NV, NY, OH, OK, PA, SC, TN, TX, WA and WI.**** International distribution: Australia and EMEA.
Product codes and lots
Item Number; Lot Number; Item Description; UDI #; Manufacturing Quantity   07.02011.001; P141352 (Z00176603); Vitality Shear-off Set Screw; (01)00889024003187 (10)P141352; 2,192 devices  07.02011.001; P141353; Vitality Shear-off Set Screw; (01)00889024003187 (10)P141353; 879 devices  07.02011.001; P141354 (Z00468651); Vitality Shear-off Set Screw; (01) 00889024003187 (10)P141354; 924 devices  07.02011.001; P141358 (Z00420148); Vitality Shear-off Set Screw; (01)00889024003187 (10)P141358; 2,193 devices  07.02011.001; P141359 (Z00274115); Vitality Shear-off Set Screw; (01)00889024003187 (10)P141359; 868 devices  07.02011.001; P141360; Vitality Shear-off Set Screw; (01)00889024003187 (10)P141360; 872 devices  07.02011.001; P142510; Vitality Shear-off Set Screw; (01)00889024003187 (10)P142510; 3,045 devices  07.02011.001; W505611; Vitality Shear-off Set Screw; (01)00889024003187 (10)W505611; 1,545 devices  07.02011.001; W508041; Vitality Shear-off Set Screw; (01)00889024003187 (10)W508041; 1,497 devices  07.02011.001; W509651; Vitality Shear-off Set Screw; (01)00889024003187 (10)W509651; 3,396 devices  07.02011.001; W509661; Vitality Shear-off Set Screw; (01)00889024003187 (10)W509661; 3,382 devices  07.02011.001; W5096619; Vitality Shear-off Set Screw; (01)00889024003187 (10)W5096619; 61 devices  07.02011.001; W526991; Vitality Shear-off Set Screw; (01)00889024003187 (10)W526991; 799 devices  07.02011.001; W529371; Vitality Shear-off Set Screw; (01)00889024003187 (10)W529371; 823 devices  07.02011.001; W530301; Vitality Shear-off Set Screw; (01)00889024003187 (10)W530301; 1,342 devices  07.02011.001; W546001; Vitality Shear-off Set Screw; (01)00889024003187 (10)W546001; 634 devices  07.02011.001; W554821; Vitality Shear-off Set Screw; (01)00889024003187 (10)W554821; 2,110 devices  07.02011.001; W559871; Vitality Shear-off Set Screw; (01)00889024003187 (10)W559871; 1,461 devices  07.02011.001; W573501; Vitality Shear-off Set Screw; (01)00889024003187 (10)W573501; 2,713 devices  07.02011.001; W573511; Vitality Shear-off Set Screw; (01)00889024003187 (10)W573511; 2,698 devices  07.02011.001; W577991; Vitality Shear-off Set Screw; (01)00889024003187 (10)W577991; 2,026 devices W508041 (01)00889024003187 (10)W508041 W509661 (01)00889024003187 (10)W509661 W526991 (01)00889024003187 (10)W526991 W530301 (01)00889024003187 (10)W530301 W554821 (01)00889024003187 (10)W554821 W573501 (01)00889024003187 (10)W573501 W577991 (01)00889024003187 (10)W577991

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0867-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0866-2019Class IITerminatedVoluntary: Firm initiated
Z-0867-2019Class IITerminatedVoluntary: Firm initiated
Z-0868-2019Class IITerminatedVoluntary: Firm initiated