Z-0177-2020Class I

Minimed Model 500 Remote Control (MMT-500, MMT-500RU, MMT-500RUS, MMT-500RSE, MMT-500RCH, MMT-500RDE...

Medtronic, Inc. / initiated 2018-08-07 / Terminated / root cause: software

Medtronic Inc. began a recall of Minimed Model 500 Remote Control (MMT-500, MMT-500RU, MMT-500RUS, MMT-500RSE, MMT-500RCH, MMT-500RDE, MMT-500REU, MMT-500RFR, MMT-500RGB, MMT-500RIL... on . FDA classified it as Class I and gives the reason as "There is a potential security vulnerability related to the use of the remote controller accessories with the insulin pumps". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0177-2020
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Minimed Model 500 Remote Control (MMT-500, MMT-500RU, MMT-500RUS, MMT-500RSE, MMT-500RCH, MMT-500RDE, MMT-500REU, MMT-500RFR, MMT-500RGB, MMT-500RIL, MMT-500RNL) for use with the MiniMed 508 Insulin Pump (MMT-508).
Product code
LZG Pump, Infusion, Insulin
Reason for recall
There is a potential security vulnerability related to the use of the remote controller accessories with the insulin pumps.
Root cause tag
software
FDA root cause
Software design
Action
Medtronic sent an Urgent Field Safety Notification letter dated August 7, 2018 to the patient, distributor, and healthcare provider via email, U.S. mail, or phone. The letter identified the affected product, problem and actions to be taken. Starting on 10/05/2021, recall notices were sent to patients, health care professionals, distributors, and personnel at clinical sites. Distributors were asked to provide their customer lists to the recalling firm so the recalling firm could send customers recall notices. The scope of the recall is being expanded to all users regardless of warranty status. Users are asked to stop using and disconnect the remote controller, disable the remote feature, and return the remote controller to the recalling firm. Customers were also asked to provide the information requested in the confirmation form to the recalling firm via phone, mail, email, or online on the firm's website. Additional questions can be directed to Technical Support at 1-800-378-2292. Disconnecting/Disabling instructions were provided in the recall notice, which provided the following link: https://info.medtronicdiabetes.com/remote-controller-return
Quantity in commerce
90,213
Distribution
Distribution was nationwide to consumers and distributors. There was no government/military distribution. Foreign distribution was made to Algeria, Australia, Austria, Belgium, Brazil, Canada, China, Croatia, Czech Republic, Denmark, Estonia, Faroe Islands, Finland, France, Germany, Greece, Hungary, Iceland, India, Iran, Iraq, Ireland, Israel, Italy, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Martinique, Namibia, Netherlands, New Caledonia, Norway, Oman, Pakistan, Poland, Reunion, Romania, Russian Federation, Saudi Arabia, Senegal, Slovakia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, and United Kingdom.
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0177-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0177-2020Class ITerminatedVoluntary: Firm initiated
Z-0178-2020Class ITerminatedVoluntary: Firm initiated