Abbott ARCHITECT c16000 Processing Module, List No. 03L77 -
Abbott Laboratories, Inc began a recall of Abbott ARCHITECT c16000 Processing Module, List No. 03L77 - Product Usage: The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The Abbott... on . FDA classified it as Class II and gives the reason as "There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1104-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Abbott ARCHITECT c16000 Processing Module, List No. 03L77 - Product Usage: The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The ARCHITECT c4000, c8000, c16000 sample pipettor assemblies include a fluid sense/pressure monitoring system that helps identify errors in aspiration.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Recall letters were issued in the U.S. 7/17/2018 via FedEx overnight mail notifying the consignee of the issue, patient impact, and necessary actions to take. Globally, primary methods of issuance were via mail, telephone, and email.
- Quantity in commerce
- 1,666 instruments
- Distribution
- Worldwide distribution - US Nationwide and countries Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belarus, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Burkina FASO, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Democratic Rep. of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Mauritania, Mauritius, Mexico, Moldova, Montenegro, Montserrat, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, St. Lucia, St. Vincent, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Venezuela, Vietnam, Yemen, Zambia, and Zimbabwe.
- Product codes and lots
- All serial numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1102-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1103-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1104-2019 | Class II | Terminated | Voluntary: Firm initiated |