Z-0980-2019Class II
Instrument Tray, Basic T2 Femur Long Catalog #: 1806-9900
Stryker Corp. / initiated 2019-01-21 / Terminated / root cause: manufacturing process
Stryker GmbH began a recall of Instrument Tray, Basic T2 Femur Long Catalog #: 1806-9900 Product Usage: Knee Arthroplasty on . FDA classified it as Class II and gives the reason as "The faceplates may incorrectly identify the T2 Basic Long Instrument Set as T2 Basic Short Instrument Set". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0980-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Instrument Tray, Basic T2 Femur Long Catalog #: 1806-9900 Product Usage: Knee Arthroplasty
- Product code
- LRO General Surgery Tray
- Reason for recall
- The faceplates may incorrectly identify the T2 Basic Long Instrument Set as T2 Basic Short Instrument Set.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Zimmer Biomet issued Urgent Field Safety Notice PFA 1982723 customer letter dated: January 21, 2019. Letter identifies affected product, problem and action to be taken: Inform individuals within your organization who need to be aware of this action. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. Response is required, even you may not have any physical inventory on site anymore. 3. Quarantine and discontinue use of the devices.
- Quantity in commerce
- 17 units
- Distribution
- In the countries of Canada and Korea
- Product codes and lots
- Lot Number(s): K026ACE (lot# packaging) KU97181 (Vendor code engraved on device)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0980-2019 | Class II | Terminated | Voluntary: Firm initiated |