Fujifilm FDR Go PLUS, Mobile X-Ray System
Fujifilm Medical Systems U.S.A., Inc. began a recall of Fujifilm FDR Go PLUS, Mobile X-Ray System Product Usage: This device is intended to be used to take a film, CR or general radiography of patients in the... on . FDA classified it as Class II and gives the reason as "FUJIFILM Medical Systems U.S.A., Inc., Modality Systems became aware of the possibility that the braking mechanism might not be engaged in all cases when the driving handle has been released". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1040-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- Fujifilm FDR Go PLUS, Mobile X-Ray System Product Usage: This device is intended to be used to take a film, CR or general radiography of patients in the hospital who cannot move and/or of outpatients in emergency. Exposure can be captured on film, CR or Digital Radiography (DR)>
- Product code
- IZL System, X-Ray, Mobile
- Reason for recall
- FUJIFILM Medical Systems U.S.A., Inc., Modality Systems became aware of the possibility that the braking mechanism might not be engaged in all cases when the driving handle has been released.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Fujifilm sent an Urgent Medical Device Correction and Removal letter dated January 16, 2019 to customers. The letter identified the affected product, problem and actions to be taken. Customers were advised to schedule a visit by our Service department to replace the handle assembly. Customers were provided with additional instructions to continue using the affected product safely until the correction was made. For questions call 201-675-8947.
- Quantity in commerce
- 56 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Serial Numbers: MQ0001281001 through MQ0001285004
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1040-2019 | Class II | Terminated | Voluntary: Firm initiated |