Z-1040-2019Class II

Fujifilm FDR Go PLUS, Mobile X-Ray System

FUJIFILM Healthcare Americas Corporation / initiated 2019-01-15 / Terminated / root cause: design

Fujifilm Medical Systems U.S.A., Inc. began a recall of Fujifilm FDR Go PLUS, Mobile X-Ray System Product Usage: This device is intended to be used to take a film, CR or general radiography of patients in the... on . FDA classified it as Class II and gives the reason as "FUJIFILM Medical Systems U.S.A., Inc., Modality Systems became aware of the possibility that the braking mechanism might not be engaged in all cases when the driving handle has been released". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1040-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Fujifilm FDR Go PLUS, Mobile X-Ray System Product Usage: This device is intended to be used to take a film, CR or general radiography of patients in the hospital who cannot move and/or of outpatients in emergency. Exposure can be captured on film, CR or Digital Radiography (DR)>
Product code
IZL System, X-Ray, Mobile
Reason for recall
FUJIFILM Medical Systems U.S.A., Inc., Modality Systems became aware of the possibility that the braking mechanism might not be engaged in all cases when the driving handle has been released.
Root cause tag
design
FDA root cause
Device Design
Action
Fujifilm sent an Urgent Medical Device Correction and Removal letter dated January 16, 2019 to customers. The letter identified the affected product, problem and actions to be taken. Customers were advised to schedule a visit by our Service department to replace the handle assembly. Customers were provided with additional instructions to continue using the affected product safely until the correction was made. For questions call 201-675-8947.
Quantity in commerce
56 units
Distribution
US Nationwide Distribution
Product codes and lots
Serial Numbers: MQ0001281001 through MQ0001285004

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1040-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1040-2019Class IITerminatedVoluntary: Firm initiated