Uroskop Omnia -fluoroscopic X-ray system Model # 10094910 The Uroskop Omnia is a solid state detector...

Siemens Medical Solutions USA, Inc. / initiated 2019-02-08 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Uroskop Omnia -fluoroscopic X-ray system Model # 10094910 The Uroskop Omnia is a solid state detector fluoroscopic X-ray system, primarily for urological... on . FDA classified it as Class II and gives the reason as "The swivel arm holding the monitors may become dislodged from the carrying arm". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0984-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Uroskop Omnia -fluoroscopic X-ray system Model # 10094910 The Uroskop Omnia is a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray diagnosis, endourology and minimal invasive urology/surgery).
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
The swivel arm holding the monitors may become dislodged from the carrying arm.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Siemens, issued an "URGENT: MEDICAL DEVICE CORRECTION "on 2/8/19 to its customers. The letter is identifying the problem, risk, actions to take and corrective action. The customers were instructed to do the following: Immediately examine your inventory and quarantine product subject to recall (if applicable). or To avoid the risk until implementation of the corrective action, it is strongly recommended not to move and position monitors above the system table. As an alternative and if clinically applicable, monitors can also be used when they are left in the parking position. In addition, if you may have further distributed this product, please identify your customers and notify them at once of this product recall. If the device has been sold and is therefore no longer in your possession, please forward this notice to the new owner. The service organization will contact and arrange a date to perform this corrective action with Updated Instruction XP006/19/S. Contact the service organization for an earlier appointment at 1-800-888-7436.
Quantity in commerce
6
Distribution
US Nationwide Distribution
Product codes and lots
Serial Numbers: 3414 3415 3416 3417 3418 3428

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0984-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0984-2019Class IITerminatedVoluntary: Firm initiated
Z-0985-2019Class IITerminatedVoluntary: Firm initiated