Z-0984-2019Class II
Uroskop Omnia -fluoroscopic X-ray system Model # 10094910 The Uroskop Omnia is a solid state detector...
Siemens Medical Solutions USA, Inc. / initiated 2019-02-08 / Terminated / root cause: design
Siemens Medical Solutions USA, Inc began a recall of Uroskop Omnia -fluoroscopic X-ray system Model # 10094910 The Uroskop Omnia is a solid state detector fluoroscopic X-ray system, primarily for urological... on . FDA classified it as Class II and gives the reason as "The swivel arm holding the monitors may become dislodged from the carrying arm". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0984-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Uroskop Omnia -fluoroscopic X-ray system Model # 10094910 The Uroskop Omnia is a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray diagnosis, endourology and minimal invasive urology/surgery).
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- The swivel arm holding the monitors may become dislodged from the carrying arm.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, Siemens, issued an "URGENT: MEDICAL DEVICE CORRECTION "on 2/8/19 to its customers. The letter is identifying the problem, risk, actions to take and corrective action. The customers were instructed to do the following: Immediately examine your inventory and quarantine product subject to recall (if applicable). or To avoid the risk until implementation of the corrective action, it is strongly recommended not to move and position monitors above the system table. As an alternative and if clinically applicable, monitors can also be used when they are left in the parking position. In addition, if you may have further distributed this product, please identify your customers and notify them at once of this product recall. If the device has been sold and is therefore no longer in your possession, please forward this notice to the new owner. The service organization will contact and arrange a date to perform this corrective action with Updated Instruction XP006/19/S. Contact the service organization for an earlier appointment at 1-800-888-7436.
- Quantity in commerce
- 6
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Serial Numbers: 3414 3415 3416 3417 3418 3428
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0984-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0985-2019 | Class II | Terminated | Voluntary: Firm initiated |