Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 -
Medica Corporation began a recall of Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Usage: For use in washing the probe of the... on . FDA classified it as Class II and gives the reason as "Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1294-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medica Corp.
- Product
- Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Usage: For use in washing the probe of the Medica EasyRA Chemistry Analyzer. The wash solution in the wedge is used to prevent interference between certain tests.
- Product code
- CHI Lipase-Esterase, Enzymatic, Photometric, Lipase
- Reason for recall
- Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Medica issued letters dated 2/8/19 via email Bulletin B1901 & B1902 to distributors and end-users on record. Letter describes problem, action to be taken: Immediately discontinue use of the Wash1 wedges (REF 10680) and discard any remaining kits/wedges. Complete the response form. Discuss possible concerns about Lipase results achieved using these lots of Wash wedges with your laboratory manager and/or clinician in order to decide on appropriate measures. Report any concerns to Medica for evaluation and review. Questions contact Medica Technical Support at (877) 777-5895 or via email at: [email protected].
- Quantity in commerce
- 87 kits
- Distribution
- Worldwide distribution - US Nationwide and countries of IRAQ, FRANCE, INDIA, GEORGIA, GEORGIA, FRANCE GUATEMALA, KENYA, IRAN, ESTONIA.
- Product codes and lots
- Lot #s 18066, 18133, 18235, 18317
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1294-2019 | Class II | Terminated | Voluntary: Firm initiated |