Z-1051-2019Class II

BD Synapsys Laboratory Solutions Catalog Number:444150

Becton Dickinson & Co / initiated 2019-02-21 / Terminated / root cause: software

Becton Dickinson & Co. began a recall of BD Synapsys Laboratory Solutions Catalog Number:444150 Product Usage: BD Synapsys is a laboratory software solution providing data management and workflow... on . FDA classified it as Class II and gives the reason as "BD Synapsys version 2.1 allowed the potential for a test order to be associated to an incorrect culture record". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1051-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Synapsys Laboratory Solutions Catalog Number:444150 Product Usage: BD Synapsys is a laboratory software solution providing data management and workflow management functionality across clinical diagnostic activities. The software integrates operational and clinical data points from laboratory hardware, with patient data received from external software systems allowing for the aggregation, evaluation, dissemination of knowledge using all data, anywhere, anytime. Currently Release 2 of BD Synapsys can connect to both Lab Automation (Kiestra) and BACTEC instruments.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
BD Synapsys version 2.1 allowed the potential for a test order to be associated to an incorrect culture record.
Root cause tag
software
FDA root cause
Software design
Action
BD notified Synapsys Customer via letter UPS on 02/21/19 identified problem and action taken. Customer sites were updated remotely with a database update on January 10, 2019. Accounts asked to complete and return the CUSTOMER RESPONSE FORM BD Synapsys. The issue will be resolved by the software update syngo CT VB10A_SP1 (Service Pack 1) as soon as the update is available. For questions contact BD Customer/Technical Support at 8006388663. For customers outside the US, contact your local BD representative or distributor.
Quantity in commerce
5 sites
Distribution
Worldwide Distribution - US: Nationwide in the states of OH, IL, MO, TN and the countries of Netherlands
Product codes and lots
BD Synapsys version 2.1  UDI: 00382904441500

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1051-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1051-2019Class IITerminatedVoluntary: Firm initiated