BD Synapsys Laboratory Solutions Catalog Number:444150
Becton Dickinson & Co. began a recall of BD Synapsys Laboratory Solutions Catalog Number:444150 Product Usage: BD Synapsys is a laboratory software solution providing data management and workflow... on . FDA classified it as Class II and gives the reason as "BD Synapsys version 2.1 allowed the potential for a test order to be associated to an incorrect culture record". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1051-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Synapsys Laboratory Solutions Catalog Number:444150 Product Usage: BD Synapsys is a laboratory software solution providing data management and workflow management functionality across clinical diagnostic activities. The software integrates operational and clinical data points from laboratory hardware, with patient data received from external software systems allowing for the aggregation, evaluation, dissemination of knowledge using all data, anywhere, anytime. Currently Release 2 of BD Synapsys can connect to both Lab Automation (Kiestra) and BACTEC instruments.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- BD Synapsys version 2.1 allowed the potential for a test order to be associated to an incorrect culture record.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- BD notified Synapsys Customer via letter UPS on 02/21/19 identified problem and action taken. Customer sites were updated remotely with a database update on January 10, 2019. Accounts asked to complete and return the CUSTOMER RESPONSE FORM BD Synapsys. The issue will be resolved by the software update syngo CT VB10A_SP1 (Service Pack 1) as soon as the update is available. For questions contact BD Customer/Technical Support at 8006388663. For customers outside the US, contact your local BD representative or distributor.
- Quantity in commerce
- 5 sites
- Distribution
- Worldwide Distribution - US: Nationwide in the states of OH, IL, MO, TN and the countries of Netherlands
- Product codes and lots
- BD Synapsys version 2.1 UDI: 00382904441500
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1051-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY