CERTAIN(R) BELLATEK(R) TIN ABUTMENT 3.4MM, Item Number IEDAN3 -
Biomet 3i, LLC began a recall of CERTAIN(R) BELLATEK(R) TIN ABUTMENT 3.4MM, Item Number IEDAN3 - Product Usage: BIOMET3i Restorative products are intended for use as an accessory to endosseous... on . FDA classified it as Class II and gives the reason as "During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2140-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ZimVie US Corp LLC
- Product
- CERTAIN(R) BELLATEK(R) TIN ABUTMENT 3.4MM, Item Number IEDAN3 - Product Usage: BIOMET3i Restorative products are intended for use as an accessory to endosseous dental implants for placement in the maxilla and mandible
- Product code
- NHA Abutment, Implant, Dental, Endosseous
- Reason for recall
- During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm disseminated the notices by email on 02/26/2019. The letter requested that the dental laboratories quarantine and return any unused product and to further notify the affected clinician referral.
- Quantity in commerce
- 3 units
- Distribution
- Nationwide and Puerto Rico, Canada, and Australia
- Product codes and lots
- Lot Numbers/UDI: 8442784-1/(01)00844868031130(10)8442784-1, 8442469-1/(01)00844868031130(10)8442469-1, 8442689-1/(01)00844868031130(10)8442689-1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2140-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2141-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2142-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2143-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2144-2019 | Class II | Terminated | Voluntary: Firm initiated |