Cardiohelp Emergency Drive, Part No. 70104.8002
Maquet Cardiovascular Us Sales, Llc began a recall of Cardiohelp Emergency Drive, Part No. 70104.8002 Product Usage: The Cardiohelp Emergency Drive is used in emergencies to manually drive the disposable if the... on . FDA classified it as Class I and gives the reason as "Upon servicing the gearwheel in the Emergency Drive the gearwheel may have been re-assembled incorrectly, leading to the inadequate operation of the Emergency Drive". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0986-2019
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Cardiohelp Emergency Drive, Part No. 70104.8002 Product Usage: The Cardiohelp Emergency Drive is used in emergencies to manually drive the disposable if the Cardiohelp-i fails.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- Upon servicing the gearwheel in the Emergency Drive the gearwheel may have been re-assembled incorrectly, leading to the inadequate operation of the Emergency Drive.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Maquet has issued an Urgent Medical Device Recall dated February 26, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that their Emergency Drives had already been inspected and replaced with a loaner Emergency Drive if found to be affected. Affected E-Drives were returned to Getinge for servicing. Upon servicing, a service technician will return the Emergency Drive to your facility. Customers were asked to complete and sign the customer response form to acknowledge receipt of the notification. If you have any questions, please contact your Maquet Cardiopulmonary (MCP)/Getinge representative or call the Maquet Cardiopulmonary (MCP)/Getinge Customer Support at +1 (888) 627-8383 (press option 2, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
- Quantity in commerce
- 10
- Distribution
- US Nationwide Distribution in the states of California, Florida, Illinois, Kentucky, Michigan, and New York.
- Product codes and lots
- Affected serial numbers: 90420359, 90421204, 90421188, 90421279, 90421228, 90421761, 90420904, 90420231, 90420455, 90421304