Z-1080-2019Class III

McKesson Consult Diagnostic Immunochemical Fecal Occult Blood Test, Part numbers 178-CK30T, 178-CT30...

Cypress Medical Products LLC / initiated 2019-02-27 / Terminated / root cause: labeling and UDI

Cypress Medical Products LLC began a recall of McKesson Consult Diagnostic Immunochemical Fecal Occult Blood Test, Part numbers 178-CK30T, 178-CT30, 178TBSO, 178-CM25, 178-CK50, 178-CM50, 178-CT50, 178CK30 on . FDA classified it as Class III and gives the reason as "Products were denied Clinical Laboratory Improvement Amendments (CLIA) waiver approval by the FDA, due to the packaging being too similar to domestically sourced products under the same branded label". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1080-2019
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cypress Medical Products LLC
Product
McKesson Consult Diagnostic Immunochemical Fecal Occult Blood Test, Part numbers 178-CK30T, 178-CT30, 178TBSO, 178-CM25, 178-CK50, 178-CM50, 178-CT50, 178CK30
Product code
KHE Reagent, Occult Blood
Reason for recall
Products were denied Clinical Laboratory Improvement Amendments (CLIA) waiver approval by the FDA, due to the packaging being too similar to domestically sourced products under the same branded label.
Root cause tag
labeling and UDI
FDA root cause
No Marketing Application
Action
On February 27, 2019, Cypress Medical Products, LLC, notified their single customer and distributor, McKesson Medical Surgical via Urgent Voluntary Medical Device Recall letter. McKesson then notified their affected distribution centers. The letter instructed the customer to check the stock for affected product. If the customer has affected product, the customer is to cease distribution and quarantine immediately. Recipients were also asked to complete the reply form that was included with the letter, even if you do not have any affected product in your possession at the distribution centers.
Quantity in commerce
1165 cases
Distribution
Shipped to distribution centers located in AL, AR, AZ, CA, CO, FL, GA, HI, IL, KY, LA, MA, MN, MO, MS, NH, NJ, NY, OH, PA, SC, TN, TX, UT, VA, WA, and WI.
Product codes and lots
All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1080-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1080-2019Class IIITerminatedVoluntary: Firm initiated