Z-1122-2019Class II

BD Blood Collection Assembly with Male Luer Lock, Catalog Number MBC6010

Becton Dickinson & Co / initiated 2019-03-07 / Terminated

Becton Dickinson & Company began a recall of BD Blood Collection Assembly with Male Luer Lock, Catalog Number MBC6010 on . FDA classified it as Class II and gives the reason as "The luer could break and cause the device to leak or break off and get stuck in the fistula needle port rendering the port inaccessible for dialysis. As a result, the patient would need to be re-cannulated with a new...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1122-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Blood Collection Assembly with Male Luer Lock, Catalog Number MBC6010
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
The luer could break and cause the device to leak or break off and get stuck in the fistula needle port rendering the port inaccessible for dialysis. As a result, the patient would need to be re-cannulated with a new fistula needle to obtain their dialysis treatment.
FDA root cause
Process change control
Action
Urgent Medical Device Recall notification letters dated 3/7/19 were distributed to customers.
Quantity in commerce
6,509,600
Distribution
US distribution to the states of FL and NJ.
Product codes and lots
Lot Numbers, UDI: 18041501, (17)210315(10)18041501, (01)50885403235915.   18042501, (17)210325(10)18042501, (01)50885403235915.   18050501, (17)210405(10)18050501, (01)50885403235915.   18051501, (17)210415(10)18051501, (01)50885403235915.   18052501, (17)210425(10)18052501, (01)50885403235915.   18060501, (17)210505(10)18060501, (01)50885403235915.   18061501, (17)210515(10)18061501, (01)50885403235915.   18062501, (17)210525(10)18062501, (01)50885403235915.   18070501, (17)210605(10)18070501, (01)50885403235915.   18072501, (17)210625(10)18072501, (01)50885403235915.   18080501, (17)210705(10)18080501, (01)50885403235915.   18081501, (17)210715(10)18081501, (01)50885403235915.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1122-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1122-2019Class IITerminatedVoluntary: Firm initiated