Z-1122-2019Class II
BD Blood Collection Assembly with Male Luer Lock, Catalog Number MBC6010
Becton Dickinson & Co / initiated 2019-03-07 / Terminated
Becton Dickinson & Company began a recall of BD Blood Collection Assembly with Male Luer Lock, Catalog Number MBC6010 on . FDA classified it as Class II and gives the reason as "The luer could break and cause the device to leak or break off and get stuck in the fistula needle port rendering the port inaccessible for dialysis. As a result, the patient would need to be re-cannulated with a new...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1122-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Blood Collection Assembly with Male Luer Lock, Catalog Number MBC6010
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- The luer could break and cause the device to leak or break off and get stuck in the fistula needle port rendering the port inaccessible for dialysis. As a result, the patient would need to be re-cannulated with a new fistula needle to obtain their dialysis treatment.
- FDA root cause
- Process change control
- Action
- Urgent Medical Device Recall notification letters dated 3/7/19 were distributed to customers.
- Quantity in commerce
- 6,509,600
- Distribution
- US distribution to the states of FL and NJ.
- Product codes and lots
- Lot Numbers, UDI: 18041501, (17)210315(10)18041501, (01)50885403235915. 18042501, (17)210325(10)18042501, (01)50885403235915. 18050501, (17)210405(10)18050501, (01)50885403235915. 18051501, (17)210415(10)18051501, (01)50885403235915. 18052501, (17)210425(10)18052501, (01)50885403235915. 18060501, (17)210505(10)18060501, (01)50885403235915. 18061501, (17)210515(10)18061501, (01)50885403235915. 18062501, (17)210525(10)18062501, (01)50885403235915. 18070501, (17)210605(10)18070501, (01)50885403235915. 18072501, (17)210625(10)18072501, (01)50885403235915. 18080501, (17)210705(10)18080501, (01)50885403235915. 18081501, (17)210715(10)18081501, (01)50885403235915.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1122-2019 | Class II | Terminated | Voluntary: Firm initiated |