Z-1069-2019Class III

Medtronic Pulmonary Vein Ablation Catheter(R) GOLD (PVAC GOLD) 990078 -

Medtronic, Inc. / initiated 2019-03-04 / Open, Classified / root cause: labeling and UDI

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) began a recall of Medtronic Pulmonary Vein Ablation Catheter(R) GOLD (PVAC GOLD) 990078 - Product Usage: The Pulmonary Vein Ablation Catheter GOLD is designed to be used for the... on . FDA classified it as Class III and gives the reason as "The firm issued an updated instructions for use to reduce the potential for phrenic nerve injury". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1069-2019
Classification
Class III
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Medtronic Pulmonary Vein Ablation Catheter(R) GOLD (PVAC GOLD) 990078 - Product Usage: The Pulmonary Vein Ablation Catheter GOLD is designed to be used for the creation of endocardial lesions (focal and linear) during cardiac ablation procedures for the treatment of symptomatic Atrial Fibrillation (AF). The PVAC GOLD is also intended to be used for cardiac electrophysiological (EP) mapping of pulmonary vein potentials, delivery of diagnostic pacing stimuli and verifying electrical isolation of the pulmonary veins post-treatment.
Product code
OAE Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Fibrillation
Reason for recall
The firm issued an updated instructions for use to reduce the potential for phrenic nerve injury.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
The firm initiated the communication on 03/04/2019 by letter. Consignees are asked to review the communication to understand that Medtronic is adding current recommendations for reducing the potential for PNI. The IFU will be updated globally, including completing submissions and obtaining regulatory approvals, where prior approvals are required. This event involves OUS consignees only.
Quantity in commerce
2917 units
Distribution
International distribution in the countries of Australia, Austria, Brazil, Canada, France, Germany, Hungary, Ireland, Israel, Italy, Latvia, Netherlands, Poland, Portugal, Spain, Switzerland, United Kingdom.
Product codes and lots
UDI 00643169522572  All lot numbers distributed in the past eighteen (18) months and all newly manufactured without the updated IFU are affected until product with the updated Instruction for Use (IFU) is available.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1069-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1069-2019Class IIIOngoingVoluntary: Firm initiated