Z-1138-2019Class II

IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the...

Topcon Healthcare, Inc. / initiated 2019-03-06 / Terminated / root cause: labeling and UDI

Topcon Medical Systems, Inc. began a recall of IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images... on . FDA classified it as Class II and gives the reason as "When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1138-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Topcon Healthcare, Inc.
Product
IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices without controlling or altering the functions and parameters of any medical devices or through computerized networks.
Product code
NFJ System, Image Management, Ophthalmic
Reason for recall
When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Topcon issued Device Correction Notice on 3/22/19 providing problem- there exists a situation in which the displayed patient name and the thumbnail image do not match and action to take.Topcon has corrected this issue in IMAGEnet 6 version 2.0.3 and will be providing a remote upgrade free of charge. Contact customer support team to schedule the upgrade at 866.922.6278 option 1 or by email at [email protected]
Quantity in commerce
450 units
Distribution
Nationwide
Product codes and lots
System Version 1.53

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1138-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1138-2019Class IITerminatedVoluntary: Firm initiated