Z-1373-2019Class II
CHOL+HDL+GLU test strips, REF 2412
Polymer Technology Systems / initiated 2019-01-16 / Terminated / root cause: design
Polymer Technology Systems, Inc. began a recall of CHOL+HDL+GLU test strips, REF 2412 on . FDA classified it as Class II and gives the reason as "Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1373-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Polymer Technology Systems
- Product
- CHOL+HDL+GLU test strips, REF 2412
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm initially notified some of their customers from January through March 2019 with a Customer Bulletin. On March 28, 2019, the firm issued an updated Urgent Medical Device Recall letter to all affected customers. The letter alerted customers to the potential for test strips to slip out of the analyzer. Actions to be taken by the Customer/User: 1. Closely follow the instructions for proper testing. " Make sure the analyzer is sitting on a flat, stable surface. Do not perform testing with the analyzer held in the hand. " After inserting the test strip(s), do not move the analyzer until testing is complete and the test strip is removed. " As stated in the instructions for use, if an unexpected test result is given, the test should be repeated. There is no need to return or destroy the product. 2. Distributors should immediately forward this notice to all customers/users who may have received this product. 3. Acknowledge receipt of this instruction by completing the attached form, even if you do not have affected product on hand. If you have any questions, please contact Karmen Mercer, Director of Global Customer Care directly, Monday through Friday, 8:00 AM to 5:00 PM, Eastern Time at +1 317-870-5610 or (U.S. toll-free) 877-570-8610.
- Quantity in commerce
- 67643 total
- Distribution
- Domestic distribution nationwide.
- Product codes and lots
- I808 I809 I811 I812 I814
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1371-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1372-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1373-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.