Z-1120-2019Class II
Gravity IV Set with 3-port Closed Stopcock Manifold Model CS42522E-07
Becton Dickinson & Co / initiated 2019-03-05 / Terminated
Becton Dickinson & Company began a recall of Gravity IV Set with 3-port Closed Stopcock Manifold Model CS42522E-07 on . FDA classified it as Class II and gives the reason as "The firm has initiated a recall due to a confirmed complaint trend for kinked tubing on the specified lots of the product". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1120-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- Gravity IV Set with 3-port Closed Stopcock Manifold Model CS42522E-07
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- The firm has initiated a recall due to a confirmed complaint trend for kinked tubing on the specified lots of the product.
- FDA root cause
- Under Investigation by firm
- Action
- On March 5, 2019, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed that product was being recalled due to reports of kinked tubing which can lead to occlusion of the sets and reduction in the rate of infusion. Please Take the Following Actions: 1. Immediately review your inventory for the specific Catalogues (Ref) and lot numbers listed above. Destroy all product subject to the recall following your institutions process for destruction. 2. Share this recall notification with all users of the product within your facility to ensure that they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material, so that BD may acknowledge your receipt of this notification and provide product replacement. If you require further assistance, please contact: Customer Advocacy Phone: 888-812-3266 Phone hours: 24 hours a day, 7 days a week Email: [email protected]
- Quantity in commerce
- 21220 total cases of 20
- Distribution
- Distributed to accounts in AL, CA, DC, FL, LA, MD, NE, TN, VA, and WA.
- Product codes and lots
- 18105594 18105595 18105596 18105597
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1120-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1121-2019 | Class II | Terminated | Voluntary: Firm initiated |